Rapid acquisition of the 3D MRI gradient impulse response function using a simple phantom measurement
Jürgen Rahmer1, Peter Mazurkewitz1, Peter Börnert1
1Philips Research, Hamburg, Germany.
Purpose:
To provide a simple tool for rapid measurement of the 3D gradient modulation transfer function (GMTF) of clinical MRI systems using a phantom. Knowledge of the transfer function is useful for gradient chain characterization, system calibration, and improvement of image reconstruction results.
Methods:
Starting from the well-established thin slice method used for phantom-based measurement of the 1D GMTF, we add phase encoding to partition the thin slices into voxels that act as localized field probes. From the signal phase evolution measured at the 3D voxel positions, the GMTF can be derived for cross and higher order spatial terms represented by spherical harmonics up to 3rd order.
Results:
Using spherical phantoms, 16 GMTFs representing all terms up to 3rd order harmonics can be determined in a scan time of <2 min. A large voxel volume of >1 mL yields high SNR, enabling signal acquisition using the system's body coil. The method is applied for improving system calibration and for characterizing the effect of additional hardware in the bore.
Conclusion:
The presented method seems well-suited for rapid measurement of the GMTF of a clinical system, as it delivers high-quality results in a short scan time.
More Related Videos
11:00Reliable Acquisition of Electroencephalography Data during Simultaneous Electroencephalography and Functional MRI
Published on: March 19, 2021
11:29Real-time Video Projection in an MRI for Characterization of Neural Correlates Associated with Mirror Therapy for Phantom Limb Pain
Published on: April 20, 2019
Related Concept Videos
Impulse Response
Kirchhoff's law forms an input signal equation, with the capacitor's current and voltage providing the output. Substituting the current and dividing by RC yields a differential equation. The output for an impulse input is the impulse...
What is an Electrochemical Gradient?
The chemical gradient relies on differences in the abundance of a substance on the outside versus the inside of a cell and flows from areas of high to low ion concentration. In contrast, the electrical gradient revolves around an...
Derivatives of Simple Functions
Impulse
Additionally, it can be shown that the...
Principle of Impulse and Moment
Impulse-Momentum Theorem
By writing Newton's second law of motion in terms of the momentum of an object and the external force acting on it, and simultaneously using the definition of the impulse vector, it can be shown that the total impulse on an object is equal to its...