Clinical Efficacy of Bilateral Deep Brain Stimulation Does Not Change After Implantable Pulse Generator Replacement
Marcel Niemann1, Gerd-Helge Schneider1, Andrea Kühn2
1Department of Neurosurgery, Charité University Medicine Berlin, Berlin, Germany.
Background:
Deep brain stimulation (DBS) is an approved therapy option for movement disorders such as Parkinson's disease (PD), essential Tremor (ET), and dystonia. While current research focuses on rechargeable implantable pulse generators (IPGs), little is known about changes of the motor functions after IPG replacement and the consequences of additionally implanted hardware.
Objective:
To assess changes of the motor functions, the therapy impedances, and the total electric energy delivered (TEED) after elective IPG replacement.
Methods:
We prospectively acquired the data of 47 patients with PD, ET, and dystonia treated with bilateral DBS. Motor functions were rated prior to and after surgery using the revised Unified Parkinson's Disease Rating Scale part III (MDS-UPDRS-III), the Fahn-Tolosa-Marin Tremor-Rating-Scale (FTM-TRS), and the Unified Dystonia Rating Scale (UDRS). Furthermore, the therapy impedances and TEED were assessed at the aforementioned times.
Results:
While preoperative motor scores were 48.32 ± 17.16 in PD, 39.71 ± 12.28 in ET, and 18.48 ± 16.30 in dystonia patients, postoperative scores were 47.84 ± 24.33, 32.86 ± 15.82, and 15.02 ± 15.17, respectively. Only in dystonia patients, motor scores significantly differed. Perioperative therapy impedance changes were 142.66 ± 105.35 Ω (Kinetra® to Activa® PC), -68.75 ± 43.05 Ω (Activa® PC to Activa® PC), and - 51.38 ± 38.75 Ω (Activa® PC to Activa® RC). Perioperative TEED changes were - 37.15 ± 38.87 μJ, 2.03 ± 35.91 μJ, and 12.39 ± 6.31 μJ in that first, second, and third group, respectively. Both the therapy impedances and TEED significantly differed between groups.
Conclusion:
Although there were no statistically significant changes in the motor functions of all patients after elective IPG replacement, the therapy impedances were significantly higher and TEED was significantly lower after IPG replacement with concurrent Pocket Adapter implantation.
Related Concept Videos
Pulse
The pulse serves as a clinical...
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls...
Self-Efficacy
Aneurysm II: Clinical Manifestations and Diagnostic Studies
Impedance Combination
Impedances and Admittance
Impedance is a measure of resistance to sinusoidal current flow in an AC circuit. Unlike their behavior in DC circuits, where inductors appear as short circuits and capacitors as open circuits, the behavior of these components in AC...