Incidence of false-positive transmissions during remote rhythm monitoring with implantable loop recorders
Muhammad R Afzal1, Julie Mease1, Tanner Koppert1
1Division of Cardiovascular Medicine, Wexner Medical Center at the Ohio State University Medical Center, Columbus, Ohio.
Background:
Implantable loop recorder (ILR) is preferred strategy for prolonged rhythm monitoring.
Objective:
The purpose of this study was to report the incidence and causes of false-positive (FP) diagnoses during remote monitoring with ILR.
Methods:
During a 4-week study period, all consecutive remote transmissions in patients with ILR (Reveal LINQ, Medtronic) implanted for atrial fibrillation (AF) surveillance, cryptogenic stroke (CS), and syncope were reviewed. A nurse specializing in device management and an electrophysiologist adjudicated all transmissions. Primary endpoint of the study was incidence of FP in patients with AF, CS, and syncope.
Results:
A total of 695 remote transmissions (scheduled downloads: 414; Alerts: 281) sent from 559 patients were adjudicated. The majority of patients had ILR for AF surveillance (n = 321), followed by CS (n =168) and syncope (n = 70) with nominal programming for rhythm diagnosis. Incidence of FP transmissions during the study period was 46%, 86%, and 71% in patients with AF, CS, and syncope, respectively. Incidence of FP transmissions was higher in patients with CS and syncope than in patients with AF (P <.001). For scheduled transmissions, primary causes of FP were signal dropout and undersensing; for alert transmissions, primary reasons for FP were premature atrial and ventricular ectopy.
Conclusion:
Incidence of FP during remote monitoring with nominal settings on this ILR was substantial, ranging from 46% to 86% depending on the indication for implantation. Adjudication of these transmissions required a considerable time commitment from electrophysiologists and device clinic personnel but would be required to avoid misdiagnosis and potential errors in clinical management.
More Related Videos
Related Concept Videos
Positive and Negative Feedback Loops
Prevalence and Incidence
Prevalence indicates the proportion of individuals in a population who have a specific disease or health...
False Memories
One primary source of false memories is misattribution, where individuals incorrectly associate external information...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Types of Reports II: Incident or Occurrence Report
Purposes:
In the healthcare industry, reports play a crucial role in documenting incidents within an agency. The primary objective of these reports is to ensure patient safety, uphold the...
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...