Development and validation of a new intraocular pressure estimate for patients with soft corneas
Kai-Jung Chen1, Ashkan Eliasy1, Riccardo Vinciguerra2
1School of Engineering, University of Liverpool, United Kingdom.
Purpose:
To introduce and clinically validate a new method of estimating intraocular pressure (IOP) in patients with keratoconus and soft corneas with the aim of significantly reducing dependence on corneal biomechanics.
Setting:
Vincieye Clinic, Milan, Italy, and Rio de Janeiro Corneal Tomography and Biomechanics Study Group, Brazil.
Design:
Retrospective case series.
Method:
This study comprised participants enrolled at two hospitals on two continents. Numerical analysis based on the finite element method was performed to simulate the effect of tonometric air pressure of the Corvis ST dynamic Scheimpflug analyzer on eye globes with wide variations in thickness, geometry, and tissue. The numerical predictions of ocular behavior were used to develop a new algorithm to produce predictions of the biomechanically corrected IOP (bIOP) in eyes with a soft cornea (bIOPs). Predictions of the bIOPs were assessed in the keratoconic clinical datasets (because on average these corneas are softer) and compared with the previously developed bIOP algorithm predictions obtained for normal healthy eyes.
Results:
The study comprised 722 eyes (722 participants). The main outcome was the absence of a significant difference in IOP between healthy eyes and keratoconic eyes when the bIOP and bIOPs algorithms were used (P > .05). There was, however, a significant difference with the uncorrected Scheimpflug analyzer IOP in both groups (P < .001). Furthermore, the bIOPs predictions were significantly less affected by corneal thickness and patient age than the Scheimpflug analyzer IOP.
Conclusion:
The bIOPs algorithm was more reliable at estimating the IOP in eyes with a soft cornea and was validated for use in eyes with keratoconus.
More Related Videos
04:59Author Spotlight: Enhancing Visual Outcomes in Cataract Surgery: A Novel Technique to Prevent Posterior Capsular Opacification Through IOL Rotation
Published on: July 7, 2023
12:09Patient-specific Modeling of the Heart: Estimation of Ventricular Fiber Orientations
Published on: January 8, 2013
Related Concept Videos
In Vitro Drug Release Testing: Overview, Development and Validation
Reliability and Validity
What are Estimates?
The estimate for the mean of a sample is denoted by ͞x, whereas the mean of the population is designated as μ. Further, parameters such...
Data Validation
Nursing assessment guides are generally based on holistic models rather than medical...
Data Validation
Key parameters for method validation include:
Estimation of k and VD of Aminoglycosides