Does prewarming of i-gel improve insertion and ventilation in anaesthetised and paralysed patients? A prospective,
Aamuktha Malyadha Reddy1, Nita Varghese1, Basavaraj Herekar1
1Department of Anaesthesiology, Kasturba Medical College, Manipal, Karnataka, India.
Context:
I-gel are supraglottic airway devices with non-inflatable gel-like cuff that is believed to mould to body temperature, to seal the airway. Hence a pre-warmed i-gel may seal faster, provide better ventilation and superior leak pressure.
Aims:
To determine if pre-warming i-gel to 40°C improves insertion and efficacy of ventilation.
Methods And Materials:
A prospective, randomised, controlled trial was done on 64 patients requiring anaesthesia with muscle relaxation for short duration. For those in group W, i-gel warmed to 40°C for 15 minutes before insertion was used, whereas for those in group C, i-gel kept at room temperature (approximately 23°C) was used. The airway sealing pressure over time, number of attempts and time taken for a successful insertion were noted.
Statistical Analysis:
Mean sealing pressure between two groups was compared using independent sample t-test. Repeated Measures ANOVA was used to analyse mean sealing pressure at 0, 15 and 30 min. P value ≤0.05 was considered statistically significant.
Results:
Sealing pressure improves over time in both the groups but the mean sealing pressure was higher in group C when compared to group W at all points of time, however this was clinically and statistically insignificant. Ease of insertion, time for successful insertion, insertion attempts, intra-operative manoeuvres were all comparable between the groups with no adverse effects.
Conclusions:
Pre-warming of i-gel to 40°C does not improve the success rate of insertion or provide a higher sealing pressure in anaesthetised and paralysed patients when compared to i-gel at room temperature.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Clinical Trials
There are four phases in a clinical trial. A phase one...
Clinical Trials: Overview
Trial and Error and Algorithm
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...