Abstract:
A 78-year-old woman presented with type IIIC and IIID endoleaks after endovascular treatment for a thoracoabdominal aortic aneurysm. Mating between the ruptured segments of superior mesenteric artery stent graft was performed. After failed attempts at relining the right renal artery stent graft, a hybrid approach was used to connect the side branch to the target vessel: a stent graft was deployed into the side branch of the main graft and it was temporarily occluded with a balloon. After opening the sac, a stent graft was deployed into the right renal artery and the two stent grafts were connected with a running suture. Follow-up examination performed 1 month postoperatively by computed tomography imaging confirmed the effectiveness of this treatment.