Gadoxetate disodium-related event during image acquisition: a prospective multi-institutional study for better MR
Marie-Luise Kromrey1,2, Masatoshi Hori3, Satoshi Goshima4
1Department of Radiology, University of Yamanashi, 1110 Shimokato, Chuo, Yamanashi, 409-3898, Japan.
Purpose:
To acknowledge the facts of gadoxetate disodium-related events in Japan and to achieve better MR practice by analyzing large cohort data with various MR parameters.
Materials And Methods:
This prospective multi-institutional study included 1993 patients (1201 men, mean age 66.4 ± 12.8 years), who received dynamic MRI with gadoxetate disodium (gadoxetate group, n = 1646) or extracellular gadolinium-based contrast agents (other-GBCAs group, n = 347) between January and November 2016. Recorded data covered adverse reactions including dyspnea, breath-hold failure during acquisition, respiratory artifacts rated with a four-point scale, and MR parameters. We compared data between the two groups in whole cohort and age-, gender-, and institution-matched subcohort using χ2 test (n = 640). Logistic regression model was used to reveal independent associates of substantial artifacts in arterial phase imaging.
Results:
Transient dyspnea rarely occurred in gadoxetate or other-GBCAs group (both < 1%). Gadoxetate group (vs other-GBCAs group) showed higher rates of breath-hold failure (whole cohort, 18.2% vs 7.7%, p < 0.001; subcohort, 17.6% vs 6.3%, p < 0.001) and substantial artifacts in arterial phase (7.2% vs 2.2%, p = 0.001; 7.4% vs 1.7%, p = 0.001). With single arterial phase protocol, substantial artifacts under gadoxetate were independently associated with age (odds ratio [OR] = 1.04, p < 0.001), hearing difficulty (OR = 2.92, p = 0.008), breath-hold practice required (OR = 1.61, p = 0.039), and short acquisition time (OR = 0.43, p = 0.005). Multiple arterial phase acquisition did not reduce the incident rate of substantial artifacts.
Conclusion:
Gadoxetate disodium was associated with breath-hold failure and substantial artifacts in arterial phase imaging, but not with dyspnea in Japan. Shorter acquisition time should be used to sustain image quality in gadoxetate disodium-enhanced arterial phase imaging.
Key Points:
• Gadoxetate disodium administration leads to breath-hold failure and substantial imaging artifacts in arterial phase MRI in Japan. • Contrast agent-induced dyspnea in arterial phase and adverse reactions are rare in Japan, without showing differences between gadoxetate disodium or other extracellular gadolinium-based contrast agents. • Shorter acquisition time significantly reduces gadoxetate-induced imaging artifacts in the arterial phase.
Related Concept Videos
Light Acquisition
Integration of Synaptic Events
Characteristics of Practical Op Amps
The ratio of differential gain to the common-mode gain is defined as the common-mode rejection ratio (CMRR). This ratio quantifies the ability of operational amplifiers (op-amps) to reject common-mode...
Theoretical Foundations of Nursing Practice
Theories provide a perspective to assess patients' conditions and organize data and methods. They also assist in analyzing and interpreting information. They represent a...
Imaging Studies VII: Vascular Imaging
Imaging Studies II: Ultrasonography