Leadless pacing with Micra TPS: A comparison between right ventricular outflow tract, mid-septal, and apical implant
Christophe Garweg1,2, Bert Vandenberk1,2, Stefaan Foulon2
1Department of Cardiovascular Sciences, University of Leuven, Leuven, Belgium.
Background:
With its steerable transcatheter delivery system, the Micra can be deployed in nonapical positions within the right ventricle, potentially allowing reduction of the paced QRS width. We sought to evaluate the safety and long-term performance of the right ventricular outflow tract (RVOT) pacing using the Micra transcatheter pacing system (TPS). We also compared the paced QRS between RVOT, mid-septal, and apical implant positions.
Methods:
All patients who underwent a Micra TPS implantation at the University Hospitals of Leuven were enrolled in this observational study. Right ventricular (RV) position of the device was assessed on per-procedural ventriculography. Paced QRS was analyzed and follow-up completed at 1 month and then every 6 months.
Results:
Among the 133 patients included (mean follow-up: 13 ± 11 months), 45 were implanted in the RVOT, 58 midseptally, and 30 at the apex. All implant procedures were successful and no pericardial effusion was encountered within the 30 days post-implant. Two major complications were reported with devices implanted at the apex. Pacing impedance was significantly higher in the RVOT compared to the mid-septal and apical position (P < .001). Pacing threshold and R-wave amplitude did not differ over time in either position. The median narrowest paced QRS duration was observed in the RVOT (142 ms) compared to mid-septal (159 ms; P < .001), and apical position (181 ms; P < .001).
Conclusion:
Implantation of the Micra TPS in the RVOT is safe and feasible. Electrical performance over time was comparable to mid-septal and apical positions. The narrowest paced QRS complexes is achieved with RVOT pacing.
Related Concept Videos
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for:
Conserved Binding Sites
Binding sites are often located in large pockets, and if their location on a protein’s surface is unknown, it can be predicted using various approaches. The energetic method computationally...
Conserved Binding Sites
Ligand Binding Sites
Protein-ligand interactions are quite specific; even though numerous potential ligands surround a cellular protein at any given time, only a particular ligand can bind to that protein. Moreover, a ligand binds only to a dedicated area on the surface of the protein, known as the...
Ligand Binding Sites
The Sense of Self: Reflected Self-Appraisal and Social Comparison