Multicenter performance evaluation of the Abbott Alinity hq hematology analyzer
Christiaan L Slim1, Brigitte A Wevers2, Martijn W H J Demmers3
1Department of Clinical Chemistry, Location Medical Center Leeuwarden, Certe Medical Diagnostics & Advice, Leeuwarden, The Netherlands.
Background Alinity hq (Abbott) is a new high-throughput hematology analyzer that exclusively employs optical principles for detecting and enumerating blood cells. It reports 29 parameters, including a six-part white blood cell (WBC) differential. The aim of this multicenter study was to evaluate the analytical and clinical performance of the Alinity hq. Methods Complete blood count (CBC) results and morphological flagging were compared to that of CELL-DYN Sapphire (Abbott) and 2 × 200-cell manual differential results, on 1473 whole-blood samples from a well-defined patient population from three different clinical laboratories in the Netherlands. In addition, within-run and within-laboratory precision, linearity, limit of quantitation, carryover and sample stability were assessed. External quality assessment samples were also evaluated. Results Data analysis demonstrated strong concordance of Alinity hq results with those of CELL-DYN Sapphire for all CBC parameters, except for basophil granulocytes. Alinity hq WBC differential showed high level of agreement with manual differential results and exhibited a better agreement with manual basophil results than CELL-DYN Sapphire. The sensitivity of the Alinity hq Blast flag was 57.6%, equal to the 57.6% sensitivity of the CELL-DYN Sapphire's Blast Alert. When considering samples with ≥5% blasts, the sensitivity of the Alinity hq Blast flag was 70.0%. Analytical performance of Alinity hq was shown to be consistent with state-of-the-art (SOTA) performance characteristics. Conclusions Alinity hq CBC measurands demonstrated good overall agreement with results obtained with CELL-DYN Sapphire, as well as manual WBC differential. The analytical and clinical performance characteristics of Alinity hq make it well suited for clinical laboratories.
Background Alinity hq (Abbott) is a new high-throughput hematology analyzer that exclusively employs optical principles for detecting and enumerating blood cells. It reports 29 parameters, including a six-part white blood cell (WBC) differential. The aim of this multicenter study was to evaluate the analytical and clinical performance of the Alinity hq. Methods Complete blood count (CBC) results and morphological flagging were compared to that of CELL-DYN Sapphire (Abbott) and 2 × 200-cell manual differential results, on 1473 whole-blood samples from a well-defined patient population from three different clinical laboratories in the Netherlands. In addition, within-run and within-laboratory precision, linearity, limit of quantitation, carryover and sample stability were assessed. External quality assessment samples were also evaluated. Results Data analysis demonstrated strong concordance of Alinity hq results with those of CELL-DYN Sapphire for all CBC parameters, except for basophil granulocytes. Alinity hq WBC differential showed high level of agreement with manual differential results and exhibited a better agreement with manual basophil results than CELL-DYN Sapphire. The sensitivity of the Alinity hq Blast flag was 57.6%, equal to the 57.6% sensitivity of the CELL-DYN Sapphire's Blast Alert. When considering samples with ≥5% blasts, the sensitivity of the Alinity hq Blast flag was 70.0%. Analytical performance of Alinity hq was shown to be consistent with state-of-the-art (SOTA) performance characteristics. Conclusions Alinity hq CBC measurands demonstrated good overall agreement with results obtained with CELL-DYN Sapphire, as well as manual WBC differential. The analytical and clinical performance characteristics of Alinity hq make it well suited for clinical laboratories.
More Related Videos
Related Concept Videos
Exercise and Muscle Performance
Endurance exercises
Endurance exercises involve running, swimming, or cycling, which require repetitive movements with low force output. When a person engages in endurance exercise, a few noticeable changes occur in their skeletal muscles. For instance, the number of capillaries...
Pathophysiology of Cardiac Performance
Mass Analyzers: Overview
Sources of Self-Esteem II: Performance Feedback
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Instrumentation