19 G aspiration needle versus 19 G core biopsy needle for endoscopic ultrasound-guided liver biopsy: a prospective
Rafael A Ching-Companioni1, David L Diehl1, Amitpal S Johal1
1Department of Gastroenterology and Nutrition, Geisinger Medical Center, Danville, Pennsylvania, United States.
Background:
Endoscopic ultrasound-guided liver biopsy (EUS-LB) is a safe and effective method for accomplishing parenchymal liver biopsy. The aim of this study was to compare a 19 G aspiration needle (FNA) with a 19 G Franseen-tip core biopsy needle (FNB) for EUS-LB.
Methods:
This was a prospective, parallel group, randomized trial comparing the tissue yields and adequacy of a 19 G FNA needle vs. a 19 G FNB needle for EUS-LB. The primary outcome was length of the longest piece of liver core specimen. Secondary outcomes were aggregate specimen length, number of complete portal triads (CPTs), and adverse events. One transgastric pass and one transduodenal pass were performed with the same needle in each patient. Specimen lengths were measured before and after histological processing.
Results:
40 patients referred for EUS-LB were randomized to either the FNA group (n = 20) or the FNB group (n = 20). Both groups had similar patient characteristics. FNB biopsies yielded longer mean (standard deviation) specimen lengths (pre-processing mean 2.09 cm [0.41] vs. mean 1.47 cm [0.46], and post-processing mean 1.78 cm [0.66] vs. mean 1.05 cm [0.42]; both P < 0.001), a longer aggregate specimen length (pre-processing mean 15.78 cm [5.19] vs. 10.89 cm [4.38]; P = 0.003), and more CPTs (mean 42.6 25 vs 18.1 [9.3]; P < 0.001) compared with the FNA needle. There were no severe adverse events or difference in adverse event rate between the two needles. Post-biopsy pain was noted in 37.5 %.
Conclusion:
EUS-LB using the FNB needle delivered longer liver biopsy specimens with more CPTs than the regular (non-core) needle.
Related Concept Videos
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
The Nucleosome Core Particle
The paradox
Nucleosomes, paradoxically, perform two opposite functions simultaneously. On the one hand, their main responsibility is to protect the delicate DNA strands from physical damage and help achieve a higher compaction ratio. While on the other hand, they must allow polymerase enzymes to access DNA...
Clinical Trials
There are four phases in a clinical trial. A phase one...
Clinical Trials: Overview
Trial and Error and Algorithm
Random Error