Measurement reproducibility of slice-interleaved T1 and T2 mapping sequences over 20 months: A single center study
Jihye Jang1,2, Long H Ngo1,3, Gabriella Captur4,5,6
1Department of Medicine (Cardiovascular Division), Beth Israel Deaconess Medical Center and Harvard Medical School, Boston, MA, United States of America.
Background:
Quantifying reproducibility of native T1 and T2 mapping over a long period (> 1 year) is necessary to assess whether changes in T1 and T2 over repeated sessions in a longitudinal study are associated with variability due to underlying tissue composition or technical confounders.
Objectives:
To carry out a single-center phantom study to 1) investigate measurement reproducibility of slice-interleaved T1 (STONE) and T2 mapping over 20 months, 2) quantify sources of variability, and 3) compare reproducibility and measurements against reference spin-echo measurements.
Methods:
MR imaging was performed on a 1.5 Tesla Philips Achieva scanner every 2-3 weeks over 20 months using the T1MES phantom. In each session, slice-interleaved T1 and T2 mapping was repeated 3 times for 5 slices, and maps were reconstructed using both 2-parameter and 3-parameter fit models. Reproducibility between sessions, and repeatability between repetitions and slices were evaluated using coefficients of variation (CV). Different sources of variability were quantified using variance decomposition analysis. The slice-interleaved measurement was compared to the spin-echo reference and MOLLI.
Results:
Slice-interleaved T1 had excellent reproducibility and repeatability with a CV < 2%. The main sources of T1 variability were temperature in 2-parameter maps, and slice in 3-parameter maps. Superior between-session reproducibility to the spin-echo T1 was shown in 2-parameter maps, and similar reproducibility in 3-parameter maps. Superior reproducibility to MOLLI T1 was also shown. Similar measurements to the spin-echo T1 were observed with linear regression slopes of 0.94-0.99, but slight underestimation. Slice-interleaved T2 showed good reproducibility and repeatability with a CV < 7%. The main source of T2 variability was slice location/orientation. Between-session reproducibility was lower than the spin-echo T2 reference and showed good measurement agreement with linear regression slopes of 0.78-1.06.
Conclusions:
Slice-interleaved T1 and T2 mapping sequences yield excellent long-term reproducibility over 20 months.
More Related Videos
07:50Global and Current Research Trends of Single-Cell Sequencing in Cancer: A Bibliometric and Visualization Study
Published on: April 18, 2025
13:12Translational Brain Mapping at the University of Rochester Medical Center: Preserving the Mind Through Personalized Brain Mapping
Published on: August 12, 2019
Related Concept Videos
Lattice Centering and Coordination Number
Types of Unit Cells
Imagine taking a large number of identical...
Center of Gravity
Center of Gravity
To determine its location, the principle of moments can be utilized by dividing the object into...
Center of Mass
The knowledge of the center of mass can also help us to describe and predict the motion of objects. For example, when a ball is thrown...
Instantaneous Center of Zero Velocity
To analyze this, consider two points on the wheel: point A and point B. The absolute velocity of point B can be expressed as the vector sum of the absolute velocity of point A and the relative velocity of point B with respect to point A. To simplify this analysis,...
Patient-centered Care