Related Experiment Video
Updated: Jan 21, 2026

Author Spotlight: A Streamlined and Accessible Analysis Method to Quantify Optokinetic Reflex Tracking Responses
Published on: April 12, 2024
A Novel Speckle-Tracking Based Method for Quantifying Tricuspid Annular Velocities in TEE
Eckhard Mauermann1, Michael Vandenheuvel2, Katrien François3
1Department for Anesthesiology and Perioperative Medicine, Ghent University Hospital, Ghent, Belgium; Department for Anesthesiology, Surgical Intensive Care, Prehospital Emergency Medicine, and Pain Therapy, Basel University Hospital, Basel, Switzerland.
Objectives:
A novel speckle-tracking-based option for measuring tricuspid annular velocities in the midesophageal 4-chamber view (ME4C) was compared with velocities measured by tissue Doppler in the apical-4 chamber view (AP4C). Because this method was based on a modified speckle-tracking-based measurement of tricuspid annular plane systolic excursion (TAPSE), the authors also compared TAPSE by speckle tracking in the ME4C with TAPSE by M-mode in the AP4C. The authors hypothesized that velocities measured by speckle tracking in transesophageal echocardiography (TEE) would be similar, correlate, and agree with those measured by tissue Doppler in transthoracic echocardiography (TTE).
Design:
Prospective diagnostic study with randomization of the order of post-induction echocardiography views by TTE (AP4C) and TEE (ME4C). Images were both acquired and analyzed by 2 echocardiographers independently. The primary outcome was S'; secondary outcomes were E', A', and TAPSE.
Setting:
Single university hospital.
Participants:
Consecutive adult patients undergoing cardiac surgery (mainly coronary artery bypass grafting).
Interventions:
None.
Main Results:
Complete data was available in 24 of 25 patients. For the primary outcome, S' measured by speckle tracking in the ME4C correlated and agreed with S' measured by tissue Doppler in the AP4C (S'STE = 0.87STDI + 0.60, p < 0.001, r = 0.78; mean bias -0.6 cm/s, 95% limits of agreement (LoA) -3.5 to 2.4 cm/s). Similarly results were found for E', but not A' (E'STE = 0.69E'TDI + 2.37, p < 0.001, r = 0.71; mean bias 0.1 cm/s, 95%LoA -2.5 to 2.8cm/s; A'STE = 0.15A'TDI + 11.17, p = 0.629). TAPSE measurements by the authors' modified speckle-tracking-based technique were similar to TAPSE by M-mode (18.2 ± 5.5 mm and 17.1 ± 3.9 mm, respectively).
Conclusions:
Tricuspid annular velocities (S'STE, E'STE) determined by speckle tracking in TEE seem to be promising surrogates for velocities measured in TTE. This may be important for perioperative assessment of the right ventricle.
Related Concept Videos
Velocity and Position by Integral Method
Consider an example to calculate the velocity and position from the acceleration function. A motorboat is traveling at a constant velocity of 5.0 m/s when it starts to decelerate to arrive at the dock. Its acceleration is...
Velocity and Position by Graphical Method
Average Velocity
Instantaneous Velocity - II
Escape Velocity
To calculate the escape velocity, it is assumed that no energy is lost to any frictional forces. In practice, a satellite...
Velocity Potential