Technical Note: Quality assurance of proton central axis pencil-beam spread-out Bragg peak using large-diameter
Peng Wang1, Mingyao Zhu1, Katja Langen1
1Department of Radiation Oncology, University of Maryland School of Medicine, Baltimore, MD, 21201, USA.
Purpose:
The aim of this work is to describe a method of machine quality assurance (QA) by measuring proton spread-out Bragg peak constructed by the integrated depth dose via a large-diameter (12-cm) multilayer ionization chamber (LD-MLIC).
Methods:
Two types of contours are used to create the nominal plan. The final nominal plan is composed of mixed-energy proton pencil-beam spots located close to the central axis. The integrated depth dose (IDD) curve contains a flat SOBP region. The LD-MLIC-measured IDD was compared to the IDD curve exported from the treatment planning system (TPS). In addition, three plans with intentionally modified energy layers to simulate wrong-delivered energy layers were created and measured by the LD-MLIC. The water equivalent thickness (WET) difference between the inserted and replaced energies was 0.2 cm. Six weeks of measurements were analyzed. A low-pass filter was introduced to mitigate the high-frequency noise in the IDD signal ratios. The filtered IDD signal ratios between the modified plans in different weeks and the baseline were used to check the energy accuracy.
Results:
The differences between the LD-MLIC-measured and TPS-exported IDDs of the nominal plan were within 2% in most parts of the curve. Bumps/dips (~1%) were noted in the filtered IDD ratio between the modified plans and the baseline.
Conclusions:
The LD-MLIC can be used to check the accuracy of multiple energies Bragg peak locations quickly in proton machine QA. The LD-MLIC was sensitive in identifying an erroneous energy with 0.2 cm in WET.
Related Concept Videos
Quality Assurance
Ionization Energy
Spreading of Chromatin Modifications
Writers
The writer...
Hypothalamic-Pituitary Axis
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Perpendicular-Axis Theorem
Consider a circular disc of mass M and radius R lying along an x-y plane. The origin lies at the center of the disc, and the z-axis is perpendicular to the disc's plane. All three axes coincide at the disc's center. The moment of inertia of this...