Feasibility of transanal total mesorectal excision (taTME) using the Medrobotics Flex® System
Heather Carmichael1, Anthony P D'Andrea2, Matthew Skancke3
1Department of Surgery, University of Colorado, Denver, CO, USA.
Background:
The use of transanal total mesorectal excision (taTME) for treatment of rectal cancer is growing, but anatomic constraints prevent access to the proximal rectum with rigid instruments. The articulated instrumentation of current surgical robots is promising in overcoming these limitations, but the bulky size of current platforms inhibits the proximal reach of dissection. Flexible robotic systems could overcome these constraints while maintaining a stable platform for dissection. The goal of this study was to evaluate feasibility of performing taTME using the semi-robotic Flex® System (Medrobotics Corp., Raynham, MA) in human cadavers.
Methods:
taTME was performed by two surgeons in six fresh human cadaveric specimens using the Flex® System, with or without transabdominal laparoscopic assistance. Both mid- and low-rectal lesions were simulated. Metrics including quality of visualization, maintenance of pneumorectum, maneuverability of instruments, effectiveness of pursestring suture placement, and dissection in an anatomically correct plane were evaluated.
Results:
The semi-robotic endoluminal platform allowed for excellent visualization, insufflation, and dissection during taTME. Adequate pursestring occlusion of the rectum was achieved in all six cases. In cadavers with simulated mid-rectal lesions (N = 4), dissection and anterior peritoneal entry was achieved in all cases, with abdominal assistance utilized in two of four cases. In cadavers with simulated low-rectal lesions (N = 2), dissection was incomplete and aborted due to difficulty maneuvering instruments in close proximity to the rigid transanal port.
Conclusions:
Use of the Flex® system for taTME is feasible for mid-rectal dissection. Advantages over the traditional multi-armed robot include longer reach of instruments with the ability to dissect up to 17 cm from the anal verge, as well as tactile feedback. The current design of the flexible platform does not permit safe dissection in the distal rectum, although this constraint may be resolved with future adjustments to the equipment.
Related Concept Videos
Base Excision Repair
The first step of...
Base Excision Repair
Nucleotide Excision Repair
Nucleotide Excision Repair
Cells are regularly exposed to mutagens—factors in the environment that can damage DNA and generate mutations. UV radiation is one of the most common mutagens and is estimated to introduce a significant number of changes in DNA. These include bends or kinks in the structure, which can block DNA replication or transcription. If these errors are not fixed, the damage can cause mutations, which in turn can result in cancer or disease depending on which sequences are...