Precision of dual-photon absorptiometry.

C C Shipp1, P S Berger, M S Deehr

  • 1USDA Human Nutrition Research Center on Aging, Tufts University, Boston, MA 02111.

Calcified Tissue International
|May 1, 1988
PubMed
Summary

Dual-photon absorptiometry (DPA) bone mineral density (BMD) measurements showed software version 08C improved accuracy after a source change, unlike version 08B. This enhances precision for spine BMD assessment in clinical practice.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Optimisation of vitamin D status in global populations.

Osteoporosis international : a journal established as result of cooperation between the European Foundation for Osteoporosis and the National Osteoporosis Foundation of the USA·2024
Same author

Previous fracture and subsequent fracture risk: a meta-analysis to update FRAX.

Osteoporosis international : a journal established as result of cooperation between the European Foundation for Osteoporosis and the National Osteoporosis Foundation of the USA·2023
Same author

Comparing Prevalence of Sarcopenia Using Twelve Sarcopenia Definitions in a Large Multinational European Population of Community-Dwelling Older Adults.

The journal of nutrition, health & aging·2023
Same author

Effects of Vitamin D with Calcium and Associations of Mean 25-Hydroxyvitamin D Levels with 3-Year Change in Muscle Performance in Healthy Older Adults in the Boston STOP IT Trial.

Calcified tissue international·2022
Same author

A Randomized Study on the Effect of Dried Fruit on Acid-Base Balance, Diet Quality, and Markers of Musculoskeletal Health in Community Dwelling Adults.

Journal of the American Nutrition Association·2022
Same author

Update of the fracture risk prediction tool FRAX: a systematic review of potential cohorts and analysis plan.

Osteoporosis international : a journal established as result of cooperation between the European Foundation for Osteoporosis and the National Osteoporosis Foundation of the USA·2022

Area of Science:

  • Medical Imaging
  • Bone Densitometry
  • Radiological Physics

Background:

  • Dual-photon absorptiometry (DPA) is a key technique for measuring bone mineral density (BMD).
  • Assessing the precision and accuracy of DPA is crucial for reliable osteoporosis diagnosis and management.
  • Software versions and source changes can potentially influence DPA measurement outcomes.

Purpose of the Study:

  • To evaluate the precision of DPA measurements in phantom and human subjects.
  • To assess the impact of a 153gadolinium source change on BMD measurements.
  • To compare the performance of two software versions (08B and 08C) in analyzing DPA data.

Main Methods:

  • DPA measurements were performed on a lumbar spine phantom and nine human subjects.

Related Experiment Videos

  • Phantom studies included varying amounts of acrylic to simulate soft tissue thickness.
  • Measurements were taken before and after a 153gadolinium source change, analyzed with Lunar Radiation Corp. software versions 08B and 08C.
  • Main Results:

    • Software version 08C corrected DPA overestimation of BMD in phantoms with thick soft tissue, unlike version 08B.
    • A significant decrease in spine BMD was observed with software 08B after the source change, but not with 08C.
    • Precision for spine BMD measurements was 1.6% with 08B and 2.0% with 08C; other regions showed higher variability.

    Conclusions:

    • The newer software version 08C demonstrates improved accuracy and precision for DPA spine BMD measurements, particularly after a source change.
    • Software version 08B showed significant bias with a cold source and a source change, impacting spine BMD accuracy.
    • These findings highlight the importance of software updates and calibration in maintaining the reliability of DPA bone densitometry.