Peak Cough Flow Fails to Detect Upper Airway Collapse During Negative Pressure Titration for Cough-Assist
Matthieu Lacombe1, Aurélien Boré1, Lorena Del Amo Castrillo1
1Réanimation Médicale, Physiologie - Explorations Fonctionnelles, et Centre d'Investigation Clinique, UMR 1429, Hôpital Raymond Poincaré, Garches, France.
Objective:
To study the ability of peak cough flow (PCF) and effective cough volume, defined as the volume exsufflated >3 L/s, to detect upper airway collapse during mechanical insufflation-exsufflation (MI-E) titration in neuromuscular patients.
Design:
Prospective observational study.
Setting:
Rehabilitation hospital.
Participants:
Patients (N=27) with neuromuscular disease causing significant impairment of chest wall and/or diaphragmatic movement.
Interventions:
The lowest insufflation pressure producing the highest inspiratory capacity was used. Exsufflation pressure was decreased from -20 cm H2O to -60/-70 cm H2O, in 10-cm H2O decrements, until upper airway collapse was detected using the reference standard of flow-volume curve analysis (after PCF, abrupt flattening or flow decrease vs previous less negative exsufflation pressure).
Main Outcome Measures:
PCF and effective cough volume profiles during expiration with MI-E.
Results:
Upper airway collapse occurred in 10 patients during titration. Effective cough volume increased with decreasing expiratory pressure then decreased upon upper airway collapse occurrence. PCF continued to increase after upper airway collapse occurrence. In 5 other patients, upper airway collapse occurred at the initial -20 cm H2O exsufflation pressure, and during titration, PCF increased and effective cough volume remained unchanged at <200 mL. PCF had 0% sensitivity for upper airway collapse, whereas effective cough volume had 100% sensitivity and specificity.
Conclusion:
Of 27 patients, 15 experienced upper airway collapse during MI-E titration. Upper airway collapse was associated with an effective cough volume decrease or plateau and with increasing PCF. Accordingly, effective cough volume, but not PCF, can detect upper airway collapse.
Related Concept Videos
Precipitation Titration: Endpoint Detection Methods
In the Volhard method, a standard excess of AgNO3 is first added to the...
Hypothesis: Accept or Fail to Reject?
There are two ways to indicate that the null hypothesis is not rejected. 'Accept' the null...
Titration of a Polyprotic Acid
Acid-Base Titration Curves
For a titration carried out for 25.00 mL of...
Titration Calculations: Weak Acid - Strong Base
For the titration of 25.00 mL of 0.100 M CH3CO2H with 0.100 M NaOH, the reaction can be represented as:
Constant Pressure Calorimetry