Are Peripherally Inserted Central Catheters Suitable for Cardiac Output Assessment With Transpulmonary
Sonia D'Arrigo1, Claudio Sandroni2, Sofia Cacciola1
1Department of Anesthesiology, Intensive Care and Emergency Medicine, Fondazione Policlinico Universitario "Agostino Gemelli" IRCCS, Rome, Italy.
Insights
Peripherally inserted central catheters (PICCs) overestimate hemodynamic measurements like cardiac index when used for transpulmonary thermodilution. This overestimation is more pronounced with double-lumen PICCs, impacting critical care monitoring accuracy.
Area of Science:
- Critical Care Medicine
- Cardiovascular Physiology
- Medical Device Technology
Background:
- Peripherally inserted central catheters (PICCs) are increasingly utilized in intensive care units (ICUs) for intravenous infusions.
- The reliability of PICCs for hemodynamic monitoring using transpulmonary thermodilution remains uncertain.
- Accurate hemodynamic assessment is crucial for managing critically ill patients.
Purpose of the Study:
- To evaluate the agreement between transpulmonary thermodilution measurements obtained via PICC and central venous catheters (CVCs).
- To assess the accuracy of hemodynamic monitoring using a Pulse Contour system (VolumeView/EV1000) calibrated with transpulmonary thermodilution via PICCs.
- To determine if PICC size and lumen configuration affect measurement reliability.
Main Methods:
- A prospective method-comparison study was conducted in a 20-bed medical-surgical ICU.
- Twenty adult ICU patients with hemodynamic instability and both PICC and CVC in place were enrolled.
- Transpulmonary thermodilution measurements were compared using bolus injections through CVCs (reference) and 4F single-lumen or 5F double-lumen PICCs, with both manual and automated injection systems.
Main Results:
- Use of a 5F double-lumen PICC led to significant overestimation of cardiac index (31% bias) compared to CVC.
- Global end-diastolic index, extravascular lung water index, and stroke volume index were also significantly overestimated with the 5F double-lumen PICC.
- A 4F single-lumen PICC showed less, but still significant, overestimation of cardiac index (12.7% bias).
- Differences persisted even with automated injection, confirming reliability issues.
Conclusions:
- Bolus injection via PICCs for transpulmonary thermodilution with the EV1000 system results in significant overestimation of key hemodynamic parameters.
- The overestimation is more severe with 5F double-lumen PICCs, questioning their suitability for accurate hemodynamic monitoring.
- Clinical decisions based on these measurements may be compromised, necessitating caution when using PICCs for this purpose.
Objectives:
Peripherally inserted central catheters are increasingly used in ICU as an alternative to centrally inserted central catheters for IV infusion. However, their reliability for hemodynamic measurements with transpulmonary thermodilution is currently unknown. We investigated the agreement between transpulmonary thermodilution measurements obtained with bolus injection through peripherally inserted central catheter and centrally inserted central catheter (reference standard) using a transpulmonary thermodilution-calibrated Pulse Contour hemodynamic monitoring system (VolumeView/EV1000).
Design:
Prospective method-comparison study.
Setting:
Twenty-bed medical-surgical ICU of a teaching hospital.
Patients:
Twenty adult ICU patients who required hemodynamic monitoring because of hemodynamic instability and had both peripherally inserted central catheter and centrally inserted central catheter in place.
Intervention:
The hemodynamic measurements obtained by transpulmonary thermodilution after injection of a cold saline bolus via both centrally inserted central catheter and either a single-lumen 4F or a double-lumen 5F peripherally inserted central catheter using were compared. In order to rule out bias related to manual injection, measurements were repeated using an automated rapid injection system.
Measurements And Main Results:
A total of 320 measurements were made. Cardiac index was significantly higher when measured with double-lumen 5F peripherally inserted central catheter than with centrally inserted central catheter (mean, 4.5 vs 3.3 L/min/m; p < 0.0001; bias, 1.24 L/min/m [0.27, 2.22 L/min/m]; bias percentage, 31%). Global end-diastolic index, extravascular lung water index, and stroke volume index were also overestimated (853 ± 240 vs 688 ± 175 mL/m, 12.2 ± 4.2 vs 9.4 ± 2.9 mL/kg, and 49.6 ± 14.9 vs 39.5 ± 9.6 mL/m, respectively; p < 0.0001). Lower, albeit significant differences were found using single-lumen 4F peripherally inserted central catheter (mean cardiac index, 4.2 vs 3.7 L/min/m; p = 0.043; bias, 0.51 L/min/m [-0.53, 1.55 L/min/m]; bias percentage, 12.7%). All differences were confirmed, even after standardization of bolus speed with automated injection.
Conclusions:
Bolus injection through peripherally inserted central catheter for transpulmonary thermodilution using EV1000 led to a significant overestimation of cardiac index, global end-diastolic index, extravascular lung water index, and stroke volume index, especially when double-lumen 5F peripherally inserted central catheter was used (ClinicalTrial.gov NCT03834675).
Related Concept Videos
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...
Exercise and Cardiac Output
Sustained exercise increases the muscles' oxygen demand, which can be...
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac...
Peripherally and Centrally Acting Muscle Relaxants: A Comparison
Centrally acting muscle relaxants can be further divided into spasmolytic and antispasmodic drugs. Spasmolytic...