Related Experiment Video
Updated: Jan 21, 2026

Author Spotlight: Assessment of Cardiac Output Calculation by Thermodilution in Pigs for Effective Perfusion Flow During EVLP
Published on: June 28, 2024
Are Peripherally Inserted Central Catheters Suitable for Cardiac Output Assessment With Transpulmonary
Sonia D'Arrigo1, Claudio Sandroni2, Sofia Cacciola1
1Department of Anesthesiology, Intensive Care and Emergency Medicine, Fondazione Policlinico Universitario "Agostino Gemelli" IRCCS, Rome, Italy.
Objectives:
Peripherally inserted central catheters are increasingly used in ICU as an alternative to centrally inserted central catheters for IV infusion. However, their reliability for hemodynamic measurements with transpulmonary thermodilution is currently unknown. We investigated the agreement between transpulmonary thermodilution measurements obtained with bolus injection through peripherally inserted central catheter and centrally inserted central catheter (reference standard) using a transpulmonary thermodilution-calibrated Pulse Contour hemodynamic monitoring system (VolumeView/EV1000).
Design:
Prospective method-comparison study.
Setting:
Twenty-bed medical-surgical ICU of a teaching hospital.
Patients:
Twenty adult ICU patients who required hemodynamic monitoring because of hemodynamic instability and had both peripherally inserted central catheter and centrally inserted central catheter in place.
Intervention:
The hemodynamic measurements obtained by transpulmonary thermodilution after injection of a cold saline bolus via both centrally inserted central catheter and either a single-lumen 4F or a double-lumen 5F peripherally inserted central catheter using were compared. In order to rule out bias related to manual injection, measurements were repeated using an automated rapid injection system.
Measurements And Main Results:
A total of 320 measurements were made. Cardiac index was significantly higher when measured with double-lumen 5F peripherally inserted central catheter than with centrally inserted central catheter (mean, 4.5 vs 3.3 L/min/m; p < 0.0001; bias, 1.24 L/min/m [0.27, 2.22 L/min/m]; bias percentage, 31%). Global end-diastolic index, extravascular lung water index, and stroke volume index were also overestimated (853 ± 240 vs 688 ± 175 mL/m, 12.2 ± 4.2 vs 9.4 ± 2.9 mL/kg, and 49.6 ± 14.9 vs 39.5 ± 9.6 mL/m, respectively; p < 0.0001). Lower, albeit significant differences were found using single-lumen 4F peripherally inserted central catheter (mean cardiac index, 4.2 vs 3.7 L/min/m; p = 0.043; bias, 0.51 L/min/m [-0.53, 1.55 L/min/m]; bias percentage, 12.7%). All differences were confirmed, even after standardization of bolus speed with automated injection.
Conclusions:
Bolus injection through peripherally inserted central catheter for transpulmonary thermodilution using EV1000 led to a significant overestimation of cardiac index, global end-diastolic index, extravascular lung water index, and stroke volume index, especially when double-lumen 5F peripherally inserted central catheter was used (ClinicalTrial.gov NCT03834675).
Related Concept Videos
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...
Exercise and Cardiac Output
Sustained exercise increases the muscles' oxygen demand, which can be...
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac...
Peripherally and Centrally Acting Muscle Relaxants: A Comparison
Centrally acting muscle relaxants can be further divided into spasmolytic and antispasmodic drugs. Spasmolytic...

