Dosimetric Effect of Rotational Setup Errors in Single-Isocenter Volumetric-Modulated Arc Therapy of Multiple Brain
Kasirajan Tamil Selvan1,2, Ganesan Padma1, Murugesan Kesavan Revathy1
1Department of Oncology, Apollo Cancer Hospital, Hyderabad, Telangana, India.
Aim:
The aim of this study is to investigate the dosimetric effects of rotational uncertainties in patient positioning on target coverage in stereotactic radiosurgery (SRS) of multiple brain metastases using single isocenter volumetric modulated arc therapy (VMAT).
Materials And Methods:
Ten SRS cases with multiple brain metastases (2-8) planned with single isocenter non-coplanar VMAT technique were investigated in this study. Pitch, yaw and roll displacements of 1, 3 and 5° were simulated about isocenter along each axes and nine plans were obtained for each case. Gross tumor volume (GTV) coverage obtained on simulated plans were compared with the original plan on four metrics of mean dose (Dmean), minimum dose to GTV (Dmin), dose to 95% of GTV (D95) and the volume covered by the 95% of the prescribed dose (V95).
Results:
At 1° rotation about 4% of the PTV had V95 and D95 values <95%. The minimum dose obtained across all PTV was 85.1% and there was not much change in the mean dose values obtained. The PTV volume which had V95 and D95 values <95% were in the range of 0.05 - 0.07cc and at a radial distance of 6.2 cm - 7.2 cm. At 3° rotation almost 50% of the PTV had V95 and D95 values <95%. The minimum dose obtained across all PTV was 48.3% and the mean dose reduced to as low as 78.8%. At 5° rotation almost 74% of the PTV had V95 and D95 values <95%. The minimum dose obtained across all PTV was 21.2% and the mean dose reduced to as low as 49.2%.
Conclusion:
Our results indicate that correcting rotational uncertainties is critical in single-isocenter, multi-target SRS. For rotational deviations, radial distance of the target from isocenter along the respective axis has a strong influence on target coverage. For rotational setup deviation at a given radial distance larger targets tend to have lesser geometric miss compared to smaller targets. Mathematical model for spherical targets can be used to estimate V95 for given rotational errors.
Related Concept Videos
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Fundamental Attribution Error
Random Error
Margin of Error
Contaminants and Errors
Another key consideration is determining the appropriate number of samples required to...
Standard Error of the Mean
The standard error of the mean is an example of a standard error. It is a unique standard deviation known as the standard deviation of the sampling distribution of the mean. The standard error of the mean is a statistic that calculates how correctly a sample distribution represents a...