Impact of apical foreshortening on deformation measurements: a report from the EACVI-ASE Strain Standardization Task
Serkan Ünlü1, Jürgen Duchenne1, Oana Mirea1
1Department of Cardiovascular Diseases, University Hospital Leuven, Herestraat 49, 3000 Leuven, Belgium.
Aims:
Foreshortening of apical views is a common problem in echocardiography. It results in an abnormally thick false apex and a shortened left ventricular (LV) long axis. We sought to evaluate the impact of foreshortened (FS) on LV ejection fraction (LVEF) and layer-specific 2D speckle tracking based segmental (S) and global (G) longitudinal strain (LS) measurements.
Methods And Results:
We examined 72 participants using a GE Vivid E9 system. FS apical views were collected from an imaging window one rib-space higher than the optimal images. Ejection fraction as well as layer-specific GLS and SLS measurements were analysed by GE EchoPAC v201 and TomTec Image Arena 4.6 and compared between optimal and FS images. On average, LV long axis was 10% shorter in FS images than in optimal images. FS induced a relative change in LVEF of 3.3% and 6.9% for GE and TomTec, respectively (both, P < 0.001). Endocardial GLS was 9.0% higher with GE and 23.2% with TomTec (P < 0.001). Midwall GLS measurements were less affected (7.8% for GE and 14.1% for TomTec, respectively, both P < 0.001). Segmental strain analysis revealed that the mid-ventricular and apical segments were more affected by foreshortening, and endocardial measurements were more affected than midwall.
Conclusion:
Optimal image geometry is crucial for accurate LV function assessment. Foreshorhening of apical views has a substantial impact on longitudinal strain measurements, predominantly in the apex and in the endocardial layer. Our data suggest that measuring midwall strain might therefore be the more robust approach for clinical routine use.
More Related Videos
Related Concept Videos
Measurement: Standard Units
Measurements of Strain
Units and Standards of Measurement
Measurements of physical quantities are...
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for:
Plastic Deformations
Plastic Deformations