Related Experiment Video
Updated: Jan 21, 2026

Bioelectric Analyses of an Osseointegrated Intelligent Implant Design System for Amputees
Published on: July 15, 2009
Preclinical Validation of a Novel Device Designed to Reduce Biofilms on Percutaneous Osseointegrated Abutments
Rachel Sammons1, Matthew East2, Melissa Grant1
1School of Dentistry, University of Birmingham, Birmingham.
Aim:
To assess the efficacy of a prototype cleaning device in removing biofilm from skin-penetrating abutments.
Background:
Adverse skin reactions around percutaneous osseointegrated implants have been linked to poor hygiene and infection. A cleaning device made from open-cell polyfoam has been developed to facilitate manual cleaning of abutments by the patient.
Methods:
Serratia biofilm was grown on clean and sterile conical bone-anchored hearing system abutments including both pre-2007 (straight sided) and post-2007 (conical-sided) designs. Eighteen were photographed and then subjected to cleaning using a toothbrush or the cleaning device soaked in water or a proprietary mouthwash. Biofilm on a further group of 10 conical abutments was recorded using a scanning electron microscope before and after cleaning with the test device soaked in water. Quantitative analysis of the efficiency of the cleaning was made using image analysis.
Results:
Removal of biofilm using a dry or wet toothbrush was not as effective as cleaning with the device. In 10 cases subjected to image analysis, approximately 90% of the biofilm was removed from the top third of the abutments, 85% from the middle third, and 48% from the basal third.
Conclusion:
The cleaning device is effective in removing most biofilm from the test abutments and represents a significant improvement in comparison with traditional methods such as a toothbrush.
More Related Videos
Related Concept Videos
Reliability and Validity
Design Example: Identifying the Locations of Monuments in the Field Using Global Positioning System Device
Biofilms
Group Design
Reducing Line Loss
With a step-up transformer at the source, the voltage is increased, thereby reducing the current in the transmission lines since power loss in...
Data Validation
Nursing assessment guides are generally based on holistic models rather than medical...