Related Experiment Video
Updated: Jul 13, 2026

The Use of Mixed Reality in Custom-Made Revision Hip Arthroplasty: A First Case Report
Published on: August 4, 2022
Mismatch 'never events' in hip and knee arthroplasty: a cohort and intervention study
A Odgaard1, M B Laursen2, K Gromov3
1Copenhagen University Hospital Herlev-Gentofte, Hellerup, Denmark.
Aims:
The aim of this study was to give estimates of the incidence of component incompatibility in hip and knee arthroplasty and to test the effect of an online, real-time compatibility check.
Materials And Methods:
Intraoperative barcode registration of arthroplasty implants was introduced in Denmark in 2013. We developed a compatibility database and, from May 2017, real-time compatibility checking was implemented and became part of the registration. We defined four classes of component incompatibility: A-I, A-II, B-I, and B-II, depending on an assessment of the level of risk to the patient (A/B), and on whether incompatibility was knowingly accepted (I/II).
Results:
A total of 26 524 arthroplasties were analyzed. From 12 307 procedures that were undertaken before implementation of the compatibility check, 21 class A incompatibilities were identified (real- or high-risk combinations; 0.17%; 95% confidence interval (CI) 0.11 to 0.26). From 5692 hip and 6615 knee procedures prior to implementation of the compatibility check, we found rates of class A-I incompatibility (real- or high-risk combinations unknowingly inserted) of 0.14% (95% CI 0.06 to 0.28) and 0.17% (95% CI 0.08 to 0.30), respectively. From 14 217 procedures after the introduction of compatibility checking (7187 hips and 7030 knees), eight class A incompatibilities (0.06%; 95% CI 0.02 to 0.11) were identified. This difference was statistically significant (p = 0.008).
Conclusion:
Our data presents validated estimates of the baseline incidence of incompatibility events for hip and knee arthroplasty procedures and shows that a significant reduction in class A incompatibility events is possible using a web-based recording system. Cite this article: Bone Joint J 2019;101-B:960-969.
More Related Videos
09:51The Transition to an Anterior-Based Muscle Sparing Approach Improves Early Postoperative Function but is Associated with a Learning Curve
Published on: September 7, 2022
07:33In Vitro Application of a Wireless Sensor in Flexion-Extension Gap Balance of Unicompartmental Knee Arthroplasty
Published on: May 5, 2023
Related Concept Videos
Mismatch Repair
Mismatch Repair
The Mutator Protein Family Plays a Key Role in DNA Mismatch Repair
The human genome has more than 3 billion base pairs of DNA per cell. Prior to cell division, that vast amount of genetic...