Comparative Analysis of Three Different Optimization Procedures for Coronary Bifurcation Provisional Stenting:
Wei Cai1,2, Lianglong Chen1,2, Linlin Zhang1,2
1Department of Cardiology, Fujian Medical University Union Hospital.
Background:
Post-dilation with kissing balloon dilation remains controversial in the 1-stent approach, but many technical improvements are possible to refine the final results. This study aimed to evaluate the results of different side-branch (SB) ostial treatments after main vessel stenting, including ostial optimization technique (OOT), simultaneous kissing balloon dilation (KBD) and single balloon dilation (SBD).
Methods:
Three different ostial side branch treatments (OOT, n = 6; KBD, n = 6; SBD, n = 6) were emulated in a synthetic bifurcated phantom using a second-generation sirolimus-eluting stent (Firebird2TM, Microport, Shanghai, China). Micro-computed tomography (micro-CT) and optical coherence tomography (OCT) were performed to assess morphologies.
Results:
Compared to the non-OOT procedures (SBD and KBD), OOT was characterized by the sequential dilation of two snuggling balloons, creating a longer valgus struts length (OOT: 2.13 ± 0.30 mm, SBD: 1.23 ± 0.34 mm, KBD: 1.11 ± 0.39 mm, p < 0.01), broader angulation between the main-branch and valgus struts axes (OOT: 42.72 ± 0.91°, SBD: 25.77 ± 7.81°, KBD: 31.78 ± 1.34°, p < 0.01), shorter neocarina length (OOT: 0.28 ± 0.31 mm, SBD: 0.64 ± 0.38 mm, KBD: 1.11 ± 0.37 mm, p < 0.01), larger SB ostial area (OOT: 6.76 ± 0.17 mm2, SBD: 4.78 ± 0.86 mm2, KBD: 5.87 ± 0.89 mm2, p < 0.01), and lower index of stent cell distortion (OOT: 6.67 ± 3.33%, SBD: 10.67 ± 4.23%, KBD: 20.00 ± 5.29%, p < 0.01). In addition, the rate of severe strut malapposition was lower with the OOT procedure compared with the non-OOT procedures (OOT: 2.22 ± 0.48%, SBD: 10.31 ± 0.66%, KBD: 6.74 ± 1.24%, p < 0.01).
Conclusions:
OOT, consisting of an initial proximal optimizing technique (POT) and sequential snuggling balloon dilation and then re-POT, significantly optimized the results of provisional bifurcation treatment. The physiological and clinical significance of our observations await further clarification.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Comparing the Survival Analysis of Two or More Groups
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...