Optimizing 3D FLAIR to detect MS lesions: pushing past factory settings for precise results
Augustin Lecler1, C Bouzad2,3, R Deschamps4
1Department of Radiology, Fondation Ophtalmologique Adolphe de Rothschild, 25 rue Manin, 75019, Paris, France. alecler@for.paris.
Background:
To assess the diagnostic value of three 3D FLAIR sequences with differing repetition-times (TR) at 3-Tesla when detecting multiple sclerosis (MS) lesions.
Methods:
In this prospective study, approved by the institutional review board, 27 patients with confirmed MS were prospectively included. One radiologist performed manual segmentations of all high-signal intensity lesions using three 3D FLAIR data sets with different TR of 4800 ms ("FLAIR4800"), 8000 ms ("FLAIR8000") and 10,000 ms ("FLAIR10,000") and two radiologists double-checked it. The main judgment criterion was the overall number of lesions; secondary objectives were the assessment of lesion location, as well as measuring contrast-to-noise ratio (CNR) and signal-to-noise ratio (SNR). A non-parametric Wilcoxon's test was used to compare the differing FLAIR.
Results:
The FLAIR8000 and FLAIR10,000 detected significantly more overall lesions per patient as compared with the FLAIR4800 [116.1 (± 61.7) (p = 0.02) and 115.8 (± 56.3) (p = 0.03) versus 99.2 (± 66.9), respectively]. The FLAIR8000 and FLAIR10,000 detected four and eight times more cortical or juxta-cortical lesions per patient as compared with FLAIR4800 [1.6 (± 2.2) (p = 0.001) and 4.1 (± 5.9) (p = 6 × 10-5) versus 0.4 (± 1.1), respectively]. CNR was significantly correlated to the TR value. It was significantly higher with FLAIR10,000 than it was with FLAIR8000 and FLAIR4800 [16.3 (± 3.5) versus 15 (± 2.4) (p = 0.01) and 12 (± 2.2) (p = 2 × 10-6), respectively] CONCLUSION: An optimized 3D FLAIR with a long TR significantly improved both overall lesion detection and CNR in MS patients as compared to a 3D FLAIR with factory settings.
Related Concept Videos
Factorial Design
Drive-Reduction Theory: Push Theory of Motivation
Uncertainty in Measurement: Accuracy and Precision
Accuracy and Precision
Setting Time of Cement
Optimal Foraging