Related Experiment Video
Updated: Jan 21, 2026

Direct Pressure Monitoring Accurately Predicts Pulmonary Vein Occlusion During Cryoballoon Ablation
Published on: February 26, 2013
Significant Radiation Dose Reduction Using a Novel Angiography Platform in Patients Undergoing Cryoballoon Pulmonary
Martin Schmidt1, Christian Tesche2,3, Alexander Hapfelmeier4
1Department of Cardiology, Kreisklinik Ebersberg, Ebersberg.
Objectives:
Cryoballoon pulmonary vein isolation (cPVI) in patients with atrial fibrillation requires fluoroscopic guidance, causing a relevant amount of radiation exposure. Strategies to reduce radiation exposure in electrophysiologic procedures and specifically cPVI are of great importance. The aim of this study was to evaluate a possible reduction of radiation dose using the novel Azurion 7 F12 x-ray system compared with its predecessor Allura FD10.
Methods:
In February 2017, the Philips Azurion angiography system was introduced, combining the Allura Clarity radiation dose reduction technology with a more powerful generator, improved image resolution, and a large screen display. In 173 patients undergoing cPVI by a single experienced operator in our institution between December 2016 and April 2018, dose area products (cGy×cm) and image quality were compared using Azurion 7 F12 or Allura FD10 angiography system.
Results:
A significant reduction in total radiation dose expressed as a dose area products of 524 (332; 821) cGy×cm on the Allura system compared with 309 (224; 432) cGy×cm on the Azurion system was observed (P<0.001). The number of imaging scenes recorded were 14.7 versus 13.9, and mean overall imaging quality scores (grading 4.85±0.4 with Azurion vs. 4.80±0.4 with Allura, P=0.38) and scores based on specific quality parameters were similar in both groups.
Conclusion:
Use of the new Azurion 7 F12 angiography system substantially reduced radiation doses compared with the previous generation reference system, Allura Clarity, without compromising imaging quality in patients undergoing cryoballoon pulmonary vein isolation.
More Related Videos
08:00Reduction of Iatrogenic Atrial Septal Defects with an Anterior and Inferior Transseptal Puncture Site when Operating the Cryoballoon Ablation Catheter
Published on: June 15, 2015
06:59Catheter-based Endovascular Angioplasty for Fibrosing Mediastinitis-associated Pulmonary Vein Stenosis
Published on: August 26, 2025
Related Concept Videos
Drug Dosing: Geriatric Patients
Drug Dosing: Obese Patients
Biological Effects of Radiation
Oxidation-Reduction Reactions
Radiation: Applications
The average...
Dose Size and Dosing Frequency: Determination Methods