Commutability of proficiency testing material containing amitriptyline and nortriptyline: A study within the
Karen Robijns1, Niels W Boone2, Rob T P Jansen3
1Association for Quality Assessment in Therapeutic Drug Monitoring and Clinical Toxicology (KKGT), Section of the Dutch Foundation for Quality Assessment in Medical Laboratories (SKML), The Hague, the Netherlands; Central Hospital Pharmacy, The Hague, the Netherlands; Haga Teaching Hospital, The Hague, the Netherlands; CAPHRI School for Public Health and Primary Care, Maastricht University, Maastricht, the Netherlands; Department of Clinical Pharmacy and Toxicology, Maastricht University Medical Centre+, Maastricht, the Netherlands.
Background:
External quality assessment schemes (EQAS) can provide important information regarding accuracy and comparability of different measurement methods if the sample matrices are composed of commutable material. The aim of this study was to assess the commutability of different matrices for the material used in an EQAS for amitriptyline and nortriptyline.
Methods:
Proficiency testing material (PTM) and patient samples containing amitriptyline and nortriptyline were prepared, collected, pooled, and distributed to participating laboratories for analysis. Low, medium and high concentrations of both drugs in liquid pooled human, lyophilized human and lyophilized bovine serum were tested in this study. The measurement deviation of the PTM results to the patient serum regression line were normalized by dividing trough the average within-laboratory SD (SDwl) derived from the results reported in the official EQAS, resulting in a relative residual. The commutability decision limit was set at 3 SDwl.
Results:
With 10 laboratories participating in this study, 45 laboratory couples were formed. All matrix types delivered several relative residuals outside the commutability decision limit. The number and the magnitude of relative residuals for both drugs were lower for liquid human sera as compared to lyophilized human and bovine sera.
Conclusions:
The PTM used for amitriptyline and nortriptyline is preferably prepared with human serum, although not all relative residuals are within the commutability decision limit.
Related Concept Videos
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Fischer Projections
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Newman Projections
The organic molecules rotate across the single bonds leading to numerous temporary three-dimensional structures of varying energy known as...
Plotting and Calibrating the Root Locus
The maximum gain occurs at the breakaway points between open-loop poles on the real axis, while the minimum gain is...
Calibration Curves: Correlation Coefficient