New detection configuration for low activity levels of PET tracers during the analysis of plasma samples
A Osovizky1, B Laster2, A A Wilson3
1Jerry J. Cohen Radiobiology Research Laboratory, Nuclear Engineering Department, Ben-Gurion University of the Negev, Israel; Centre for Addiction and Mental Health, Toronto, Canada; Rotem Industries Ltd, Beer-Sheva, Israel.
Abstract:
A new radio-HPLC detection system for measuring radioactivity in plasma samples during Positron Emission Tomography [PET] studies was developed. It is based on detecting both the positron and one of the annihilation photons. The system focused on improving the measurement of radioactivity concentrations on an unmetabolized positron emitting a radiopharmaceutical [PER] in the presence of its radioactive metabolites, all containing the same positron emitter. This paper presents a new detection configuration that improves the minimal detectible activity (MDA), simplify the measuring systems and reduces the error caused by the metabolites. The detector is based on a plastic scintillator and a BGO scintillation crystal, that produces different light output spectra for signal and noise events. By summing the positron and the annihilated photon light outputs, different spectra are obtained for the metabolite and for the parent compound tracer and for tracer marked by different positron emitting isotopes. This new detection system can improve quantitative analysis of plasma samples. The spectrum change provides up to a three-fold improvement in sensitivity compared to the currently used detection systems that measure only the annihilation coincidence events. Results showed that for 11C the MDA was improved by approximately 520%. Furthermore, it provides the additional advantage of reliability by providing a method for separating the signal and noise readings from the gross detector readout. Accurate reconstruction algorithm of the signal was achieved over a wide measuring range even when the signal was only 5% of the gross measurement.
More Related Videos
Related Concept Videos
Electron Configurations
The relative energies of the subshells determine the order in which atomic orbitals are filled (1s, 2s, 2p, 3s, 3p,...
Controller Configurations
Control-system compensation involves various configurations, most commonly series or cascade compensation, in which the controller...
Electron Configuration of Multielectron Atoms
Stability of Equilibrium Configuration
A stable equilibrium occurs when a system tends to return to its original position when given a small displacement, and the potential energy is at its minimum. An example of a stable equilibrium is when a cantilever beam is fixed at one end and a weight is attached to the other end. If the weight...
Configurations of BJT
The common base configuration is noted for its high voltage gain, positioning it as an ideal choice for single-stage amplifier circuits, such as microphone pre-amplifiers. A notable...
Factors Affecting Renal Clearance: Drug's Physicochemical Properties and Plasma Levels
One important factor is the drug's molecular size. The kidneys readily excrete smaller molecules below 300 Daltons (Da). On the other hand, molecules weighing between 300 and 500 Da are excreted through both urine and bile. Larger molecules above 500 Da tend to be excreted...