Related Experiment Video
Updated: Oct 2, 2026

Expired CO2 Measurement in Intubated or Spontaneously Breathing Patients from the Emergency Department
Published on: January 29, 2011
Ventilatory frequency influences accuracy of end-tidal CO2 measurements. Analysis of seven capnometers
1Department of Anesthesia, University of Iowa College of Medicine, Iowa City 52242.
Abstract:
An accurate high-frequency response is mandatory when end-tidal CO2 (PETCO2) is monitored during pediatric general anesthesia. The purpose of this study was to assess the accuracy of six infrared-based capnometers and one multiplexed mass spectrometer available at our institution at increasing frequency. Capnometers studied were the Datascope Accucap, Hewlett-Packard 47210A capnometer, Narkomed 3 Capnomed, Novametrix Capnogard model 1250, Perkin-Elmer Advantage, Puritan-Bennett Datex CO2 monitor, and Traverse Medical Monitor model 2200 capnometer. Changes in CO2 concentration were generated by a solenoid valve switching between 100% O2 and 7% CO2 in O2. Frequencies, 8-101 cycles/min were chosen to stimulate a range that might be generated by children during general endotracheal anesthesia. At every rate the displayed PETCO2 was recorded. Differences in displayed PETCO2 from known CO2 ranged from -16.4 to +6.6. At or below frequencies of 31 cycles/min, four capnometers overreported and three underreported PETCO2. At frequencies above 31 cycles/min, six capnometers underreported and one overreported PETCO2. Errors may be clinically significant if they influence ventilator settings for patients.
More Related Videos
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Respiratory Volumes and Capacities