Technical note: Glucose concentration in dairy cows measured using 6 handheld meters designed for human use
R B Lopes1, A Valldecabres1, N Silva-Del-Río2
1Veterinary Medicine Teaching and Research Center, 18830 Road 112, Tulare, CA 93274.
Six handheld glucose meters were tested for accuracy in dairy cows. While some showed good precision, none accurately measured blood glucose, indicating a need for cattle-specific algorithms.
Area of Science:
- Veterinary Medicine
- Animal Science
- Clinical Diagnostics
Background:
- Accurate blood glucose monitoring is crucial for dairy cow health management.
- Existing handheld glucose meters are designed for human use, raising questions about their applicability in cattle.
Purpose of the Study:
- To evaluate the precision and accuracy of six human handheld glucose meters for measuring blood glucose in dairy cows.
- To compare the performance of these meters against a laboratory reference method.
Main Methods:
- Blood samples from 97 Jersey and crossbreed cows were analyzed using six handheld glucose meters (Accu-Chek Aviva Plus, Aga Matrix, Contour Next, FreeStyle Precision Neo, Nova Max Plus, Precision Xtra).
- Triplicate measurements were performed for each meter.
- Plasma glucose was quantified using the hexokinase glucose-6-phosphate dehydrogenase laboratory reference method.
Main Results:
- Precision varied among meters, with Accu-Chek Aviva Plus showing the lowest coefficient of variation (2.2%).
- Lin's concordance correlation coefficients ranged from 0.48 (Aga Matrix) to 0.75 (FreeStyle Precision Neo).
- All meters exhibited bias, with wide limits of agreement; bias constituted a significant portion of the total observed error, ranging from 56.2% to 81.5%.
Conclusions:
- Despite satisfactory precision in some meters (Accu-Chek Aviva Plus, Contour Next, Precision Xtra), none demonstrated sufficient accuracy for measuring blood glucose in dairy cows.
- Significant bias and wide limits of agreement were observed across all evaluated devices.
- Future research should explore the development of cattle-specific algorithms to enhance the accuracy and precision of handheld glucose meters for veterinary use.
More Related Videos
10:35Extracellular Glucose Depletion as an Indirect Measure of Glucose Uptake in Cells and Tissues Ex Vivo
Published on: April 6, 2022
07:04Measuring Uptake of the Glucose Analog, 6-(N-(7-Nitrobenz-2-Oxa-1,3-Diazol-4-yl)Amino)-6-Deoxyglucose, in Intact Murine Neural Retina
Published on: March 14, 2025
Related Concept Videos
Drug Concentrations: Measurements
Plasma...
Concentration Cells
Consider the following voltaic cell:
Design Example: Measuring Distance Between Two Points with Obstructions
Group Design
Glucose Homeostasis: Regulation of Blood Glucose
During fasting, when blood glucose levels are low, the pancreas secretes glucagon. it...
Glucose Transporters
Facilitated diffusion-glucose transporters (GLUTs) are encoded by the solute-linked carrier (SLC) family 2, subfamily A gene family, or SLC2A. The 14 GLUT protein members are distributed into three classes: