Related Experiment Video
Updated: Jan 21, 2026

Determining the Ice-binding Planes of Antifreeze Proteins by Fluorescence-based Ice Plane Affinity
Published on: January 15, 2014
Determination of single cryoablation outcome within 30 to 60 seconds of freezing based on ice impedance
Boaz Avitall1, Ken S Lizama1, Arthur Kalinski1
1Division of Cardiology, University of Illinois at Chicago, Chicago, Illinois.
Background:
A direct indicator of effective pulmonary vein isolation (PVI) based on early ice formation is presently lacking.
Objective:
The initial impedance rise within 30 to 60 seconds (sec) of single cryoablation relating to ice on the distal surface of the cryoballoon could; predict effective PVI with early termination, the need for prolonging the cryoablation, or failure to achieve effective ablation.
Methods:
Impedance measurements were taken between two ring electrodes, at the anterior balloon surface and at the shaft behind the balloon. Ice covering the anterior ring leads to impedance rise. Single cryoablation (eight animals, 37 veins) was applied for 90 to 180 sec. Cryoapplication was terminated if the impedance reached ≥500 Ω. Impedance levels at ≤60 sec of cryoablation were divided into three groups based on the characteristics of the impedance rise. PVI was confirmed acutely and at 45 ± 9 days recovery by electrophysiology mapping and histopathology.
Results:
At 60 sec of freezing, an impedance rise of 34.1 ± 15.2 Ω (13-50 Ω) and slope of the impedance rise (measured during 15-30 sec of cryoapplication) less than 1 Ω/sec resulted in failed PVI. An impedance rise of 104.4 ± 31.5 Ω (76-159 Ω) and slope of 2 Ω/sec resulted in 100% PVIs. An impedance rise of 130.9 ± 137.8 Ω (40-590 Ω) and slope of 10 Ω/sec resulted in 100% PVIs with early termination at 90 sec.
Conclusion:
The efficacy of single cryoablation can be defined within 30 to 60 sec based on ice impedance. Three unique impedance profiles described in this investigation are associated with the uniformity and thickness of the ice buildup on the anterior surface of the balloon. One cryoablation with an adequate impedance rise is needed for successful outcomes.
More Related Videos
Related Concept Videos
Phase Transitions: Melting and Freezing
Freezing Point Depression and Boiling Point Elevation
The boiling point of a liquid is the temperature at which its vapor pressure is equal to ambient atmospheric pressure. Since the vapor pressure of a solution is lowered due to the presence of nonvolatile solutes, it stands to reason that the solution’s boiling point will subsequently be increased. Vapor pressure increases with temperature, and so a solution will require a higher temperature than will pure solvent to achieve any given vapor pressure, including one...
Predicting Reaction Outcomes
Impedance Combination
Impedances and Admittance
Impedance is a measure of resistance to sinusoidal current flow in an AC circuit. Unlike their behavior in DC circuits, where inductors appear as short circuits and capacitors as open circuits, the behavior of these components in AC...
Outcomes of Glycolysis
Cellular respiration can occur aerobically (with oxygen) or anaerobically (without oxygen). In the presence of oxygen, cellular respiration starts with glycolysis and continues with pyruvate...