The usefulness of large sample size patient dose audits for optimisation of CT automatic exposure control (AEC)
Craig S Moore1, Tim J Wood1,2, John R Saunderson1
1Medical Physics Service, Queen's Centre, Castle Hill Hospital, Hull University Teaching Hospitals NHS Trust, Castle Road, Hull, HU16 5JQ, United Kingdom.
Abstract:
The aim of this study was to demonstrate the usefulness of large sample size patient dose audits for optimisation of CT automatic exposure control (AEC) settings, even when the investigation is limited to only three scanners at a single institution. Pre-optimisation patient dose audits of common CT examinations (n > 200 for each protocol) on three CT scanners (two Philips Brilliance and one Toshiba Aquilion) using radiology information system (RIS) data were conducted showing sub-optimal CT AEC performance on the Toshiba scanner. Based on these results, an optimisation exercise was carried out on the non-optimally performing scanner by phantom measurement and investigation of system configuration. Post-optimisation patient dose audits were subsequently carried out to assess the success of the optimisation exercise demonstrating standardisation of doses; median dose-length-product values were reduced by up to 43% on the sub-optimal scanner without any adverse effect on clinical image quality. This study has demonstrated that large sample patient dose audits using RIS data can be instrumental in identifying and rectifying sub-optimal CT AEC performance, even when the investigation is limited to only three scanners at a single institution.
Related Concept Videos
Dose Size and Dosing Frequency: Determination Methods
Drug Dosing: Geriatric Patients
Drug Dosing: Obese Patients
Sample Size Calculation
The sample size for the given experiment or sampling effort is fundamental to any study design. Sample size decides the number of...
One-Way ANOVA: Equal Sample Sizes
Different sample means can result in different values for the variance estimate: variance between samples. This is because the variance between samples is calculated as the product of the sample size and the variance between the...
One-Way ANOVA: Unequal Sample Sizes