Peripheral nerve stimulation limits of a high amplitude and slew rate magnetic field gradient coil for neuroimaging
Ek T Tan1, Yihe Hua1, Eric W Fiveland1
1GE Research, Niskayuna, New York.
Purpose:
To establish peripheral nerve stimulation (PNS) thresholds for an ultra-high performance magnetic field gradient subsystem (simultaneous 200-mT/m gradient amplitude and 500-T/m/s gradient slew rate; 1 MVA per axis [MAGNUS]) designed for neuroimaging with asymmetric transverse gradients and 42-cm inner diameter, and to determine PNS threshold dependencies on gender, age, patient positioning within the gradient subsystem, and anatomical landmarks.
Methods:
The MAGNUS head gradient was installed in a whole-body 3T scanner with a custom 16-rung bird-cage transmit/receive RF coil compatible with phased-array receiver brain coils. Twenty adult subjects (10 male, mean ± SD age = 40.4 ± 11.1 years) underwent the imaging and PNS study. The tests were repeated by displacing subject positions by 2-4 cm in the superior-inferior and anterior-posterior directions.
Results:
The x-axis (left-right) yielded mostly facial stimulation, with mean ΔGmin = 111 ± 6 mT/m, chronaxie = 766 ± 76 µsec. The z-axis (superior-inferior) yielded mostly chest/shoulder stimulation (123 ± 7 mT/m, 620 ± 62 µsec). Y-axis (anterior-posterior) stimulation was negligible. X-axis and z-axis thresholds tended to increase with age, and there was negligible dependency with gender. Translation in the inferior and posterior directions tended to increase the x-axis and z-axis thresholds, respectively. Electric field simulations showed good agreement with the PNS results. Imaging at MAGNUS gradient performance with increased PNS threshold provided a 35% reduction in noise-to-diffusion contrast as compared with whole-body performance (80 mT/m gradient amplitude, 200 T/m/sec gradient slew rate).
Conclusion:
The PNS threshold of MAGNUS is significantly higher than that for whole-body gradients, which allows for diffusion gradients with short rise times (under 1 msec), important for interrogating brain microstructure length scales.
Related Concept Videos
Magnetic Fields
A magnetic field is defined by the force that a charged particle experiences...
Magnetic Field of a Solenoid
Consider a solenoid with 100 turns wrapped around a cylinder of...
Magnetic Field Lines
Magnetic field lines follow several hard-and-fast rules:
Energy In A Magnetic Field
Take an ideal inductor with zero resistance. Although it's practically impossible, assume that the coil's resistance is so small that it is practically negligible. The loss of the field's energy to dissipate thermal energy (or heat) is thus...
Peripheral Nervous System: Ganglia and Nerves
Nerves
The nerve is a bundle of axons that serves as the communication highway in the PNS. Each nerve is ensheathed in a protective layer of connective tissue called the epineurium. This outermost layer safeguards the nerve and supports the...
Magnetic Field Of A Current Loop