Real-World Data from the First U.S. Commercial Users of an Implantable Continuous Glucose Sensor
Patricia Sanchez1, Samanwoy Ghosh-Dastidar1, Katherine S Tweden1
1Senseonics, Incorporated, Germantown, Maryland.
Abstract:
The Eversense® Continuous Glucose Monitoring (CGM) System, with the first 90-day implantable sensor, received FDA (Food and Drug Administration) approval in June 2018. No real-world experience has been published. Deidentified sensor glucose (SG) data from August 1, 2018 to May 11, 2019 in the Eversense Data Management System (DMS) were analyzed for the first 205 patients who reached a 90-day wear period on the Eversense CGM system. The mean SG, standard deviation (SD), median interquartile range, coefficient of variation (CV), glucose measurement index (GMI), and percent and time in minutes across glucose ranges were computed for the 24-h time period, the nighttime (00:00-06:00), and by 30-day wear periods. Sensor accuracy, sensor reinsertion rate, transmitter wear time, and safety data were assessed. Of the 205 patients, 129 identified as type 1, 18 as type 2, and 58 were unreported. Fifty were CGM naive, 112 had prior CGM experience, and 43 were unreported. The mean SG was 161.8 mg/dL, SD was 57.4 mg/dL, CV was 0.35, and GMI was 7.18%. Percent SG at <54 mg/dL was 1.2% (18 min), <70 mg/dL was 4.1% (59.7 min), time in range (≥70-180 mg/dL) was 62.3% (897.7 min), >180-250 mg/dL was 21.9% (315.8 min), and >250 mg/dL was 11.6% (166.7 min). Nighttime values were similar. The glucometric values were similar over 30-day time periods of the sensor wear. The mean absolute relative difference (SD) using 27,708 calibration paired points against home blood glucose meters was 11.2% (11.3%). The sensor reinsertion rate was 78.5%. The median transmitter wear time was 83.6%. There were no related serious adverse events. The Eversense real-world data showed promising glycemic results, sensor accuracy, and safety. These data suggest that the Eversense CGM system is a valuable tool for diabetes management.
More Related Videos
08:25Construction of a Wireless-Enabled Endoscopically Implantable Sensor for pH Monitoring with Zero-Bias Schottky Diode-based Receiver
Published on: August 27, 2021
08:32Studying the Hypothalamic Insulin Signal to Peripheral Glucose Intolerance with a Continuous Drug Infusion System into the Mouse Brain
Published on: January 4, 2018
Related Concept Videos
Glucose Homeostasis: Regulation of Blood Glucose
During fasting, when blood glucose levels are low, the pancreas secretes glucagon. it...
Glucose Transporters
Facilitated diffusion-glucose transporters (GLUTs) are encoded by the solute-linked carrier (SLC) family 2, subfamily A gene family, or SLC2A. The 14 GLUT protein members are distributed into three classes:
Continuing Care
Continuity Equation
The mass flow rate is expressed as:
Continuity Equation
Glucose Absorption Into the Small Intestine