Purpose:
Virtual monoenergetic images (VMIs) derived from dual-energy computed tomography (DECT) have been explored for several clinical applications in recent years. However, VMIs at low and high keVs have high levels of noise. The aim of this study was to reduce image noise in VMIs by using a two-step noise reduction technique.
Methods:
VMI was first denoised using a modified highly constrained backprojection (HYPR) method. After the first-step denoising, a general-threshold filtering method was performed. Two sets of anthropomorphic phantoms were scanned with a clinical dual-source DECT system. DECT data (80/140Sn kV) were reconstructed as VMI series at 12 different energy levels (range, 40-150 keV, interval, 10 keV). For comparison, the averaged VMIs obtained from 10 repeated DECT scans were used as the reference standard. The signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR) and root-mean-square error (RMSE) were used to evaluate the quality of VMIs.
Results:
Compared to the original HYPR method, the proposed two-step image denoising method could provide better performance in terms of SNR, CNR, and RMSE. In addition, the proposed method could achieve effective noise reduction while preserving edges and small structures, especially for low-keV VMIs.
Conclusion:
The proposed two-step image denoising method is a feasible method for reducing noise in VMIs obtained from a clinical DECT scanner. The proposed method can also reduce edge blurring and the loss of intensity in small lesions.
More Related Videos
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Oxidation-Reduction Reactions
What is Energy?
Virtual Work
In static equilibrium, a body can experience an imaginary or virtual movement, such as displacement or rotation. The virtual work done by a force is equal to the dot product of force and virtual displacement in the direction of the force. When it comes to virtually rotating a...
Free Energy