Related Experiment Video
Updated: Jan 21, 2026

Author Spotlight: Studying Host-Microbe Interactions in Wound Biofilm Formation
Published on: June 16, 2023
Invisible hemolysis in serum samples interferes in NSE measurement
Antonio Mastroianni1, Rossella Panella1, Daniele Morelli1
1Laboratory Medicine Unit, IRCCS Istituto Nazionale per lo Studio e la Cura dei Tumori, Milano, Italy.
Introduction:
The accuracy of serum neuron-specific enolase (NSE) measurements is critical, particularly in neurologic diseases and cancer. NSE measurements are compromised by slight, even invisible, hemolysis, which can produce apparently higher NSE levels, leading to inappropriate clinical decisions. In this article, we describe this issue and propose a solution for avoiding incorrect results.
Methods:
Twenty blood samples from donors with NSE values that were within the reference interval were considered. Experimental hemolysis was induced in vitro to examine the relationship between the degree of hemolysis and the increase in serum NSE. The data were then subjected to statistical analysis.
Results:
There was excellent correlation (r2 0.953) between the degree of hemolysis and the rise in NSE concentration. Each hemolysis unit (equal to 1 mg/dL of free hemoglobin) corresponded to a mean value of 0.29 ± 0.09 ng/mL NSE that was released from red blood cells.
Conclusion:
The hemolysis index must be measured in every sample with no evident hemolysis before assaying it for NSE. Moreover, if the degree of hemolysis is between 5 and 30 units, the increase in NSE (from 1.5 to 9.0 ng/mL) must be calculated, and the laboratory results should be appended with comments that suggest the approximate rise in NSE.
Related Concept Videos
Interference and Diffraction
RNA Interference
This process occurs naturally in cells, often through the activity of genomically-encoded microRNAs. Researchers can take advantage of this mechanism by introducing synthetic RNAs to deactivate specific genes for research or therapeutic purposes. For example, RNAi could be used...
Interference and Decay
Interference occurs when competing memories hinder the retrieval of particular information. It can be classified into two types: proactive and retroactive interference. Proactive...
Interference: Path Lengths
Two special sources may be considered when they are in phase. This can be easily achieved by feeding the two sources from the same source. An example would be synchronizing the two speakers by feeding them with the same source, such as the sound waves produced by a tuning fork. This setup ensures that the two sources have the same frequency and are...
Interference and Superposition of Waves
Interference occurs in mechanical waves, such as sound waves, waves on a string, and surface water waves. Mechanical waves correspond to the physical displacement of particles. Hence,...
Sound Waves: Interference