Localizing Function-specific Targets for Transcranial Magnetic Stimulation in the Absence of Navigation Equipment
Published on: May 23, 2025
A new device to improve target localization for transcranial magnetic stimulation therapy
Nicholas T Trapp1, Brandt Uitermarkt2, Marcie King Johnson3
1Department of Psychiatry, University of Iowa, Iowa City, IA, United States.
Background:
Accurate identification of cranial midline structures is essential for many targeting techniques that use repetitive transcranial magnetic stimulation (rTMS), including the Beam F3 method used for depression treatment.
Objective:
Evaluate whether a novel, laser-sighted device will assist with more accurate identification of the cranial midline relative to standard scalp-based measurement procedures.
Methods:
Three trained TMS technicians performed repeated scalp-based measurements to identify the inion and vertex on five subjects (n = 54 measurements). Measurements were compared to points identified with the midline localizer device and the true midline as defined by MRI midline structures.
Results:
Use of the midline localizer was more accurate for midline identification than technician measurement (p = 0.00025) and the ratio of localizing the midline within 5 mm was higher (78% versus 54%, p = 0.008).
Conclusion:
Use of a laser-sighted midline localizer device can improve the accuracy of scalp measurements associated with target localization for rTMS treatment protocols.
More Related Videos
10:15Utilizing Repetitive Transcranial Magnetic Stimulation to Improve Language Function in Stroke Patients with Chronic Non-fluent Aphasia
Published on: July 2, 2013
10:54Conventional Repetitive Transcranial Magnetic Stimulation for Depression: A Step-by-Step Protocol
Published on: November 21, 2025
Related Concept Videos
Targeted Cancer Therapies
There are several types of targeted therapies against...
Gene Therapy
Colors and Magnetism
When atoms or molecules absorb light at the proper frequency, their electrons are excited to higher-energy orbitals. For many main group atoms and molecules, the absorbed photons are in the ultraviolet range of the electromagnetic spectrum, which cannot be detected by the human eye. For coordination compounds, the energy difference between the d orbitals often allows photons in the visible range to be absorbed and emitted, which is seen as colors by the human...
Group Therapy
Improving Translational Accuracy
Improving Translational Accuracy