Effect of micro-computed tomography reconstruction protocols on bone fractal dimension analysis
Related Experiment Video
Updated: Jan 21, 2026

Assessment of Bone Fracture Healing Using Micro-Computed Tomography
Published on: December 9, 2022
Hugo Gaêta-Araujo1, Nicolly Oliveira-Santos1, Danieli Moura Brasil1
1Department of Oral Diagnosis, Division of Oral Radiology, Piracicaba Dental School, University of Campinas (UNICAMP), Av. Limeira, Piracicaba, São Paulo, Brazil.
Objectives:
To evaluate the influence of the level of three micro-CT reconstruction tools: beam-hardening correction (BHC), smoothing filter (SF), and ring artefact correction (RAC) on the fractal dimension (FD) analysis of trabecular bone.
Methods:
Five Wistar rats' maxillae were individually scanned in a SkyScan 1174 micro-CT device, under the following settings: 50 kV, 800 µA, 10.2 µm voxel size, 0.5 mm Al filter, rotation step 0.5°, two frames average, 180° rotation and scan time of 35 min. The raw images were reconstructed under the standard protocol (SP) recommended by the manufacturer, a protocol without any artefact correction tools (P0) and 35 additional protocols with different combinations of SF, RAC and BHC levels. The same volume of interest was established in all reconstructions for each maxilla and the FD was calculated using the Kolmogorov (box counting) method. One-way ANOVA with Dunnet's post-hoc test was used to compare the FD of each reconstruction protocol (P0-P35) with the SP (α = 5%). Multiple linear regression verified the dependency of reconstruction tools in FD.
Results:
Overall, FD values are not dependent on RAC (p = 0.965), but increased significantly when the level of BHC and SF increased (p < 0.001). FD values from protocols with BHC at 45% combined with SF of 2, and BHC at 30% combined with SF of 4 or 6 had no statistical difference compared to SP.
Conclusions:
BHC and SF tools affect the FD values of micro-CT images of the trabecular bone. Therefore, these reconstruction parameters should be standardized when the FD is analyzed.
More Related Videos
04:24Proper Positioning and Restraint of a Rat Hind Limb for Focused High Resolution Imaging of Bone Micro-architecture Using In Vivo Micro-computed Tomography
Published on: November 22, 2017
08:39Longitudinal Micro-Computed Tomography Image Analysis for User-Defined Region of Interest in Critical-Sized Bone Defects
Published on: June 24, 2025
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Support Reactions in Three Dimensions
Ball and Socket Joint is one of the supports allowing free rotation about any axis. This freedom of rotation is...
Relative Velocity in One Dimension
Relative Velocity in Two Dimensions