Pre-treatment 2D and 3D dosimetric verification of volumetric arc therapy. A correlation study between gamma index
Lukasz Szczurek1,2, Robert Juszkat3, Jolanta Szczurek4,5
11st Department of Medicine, Poznan University of Medical Sciences, Poznan, Poland.
Objectives:
To evaluate methods for the pre-treatment verification of volumetric modulated arc therapy (VMAT) based on the percentage gamma passing rate (%GP) and its correlation and sensitivity with percentage dosimetric errors (%DE).
Methods:
A total of 25 patients with prostate cancer and 15 with endometrial cancer were analysed. The %GP values of 2D and 3D verifications with different acceptance criteria (1%/1 mm, 2%/2 mm, and 3%/3 mm) were obtained using OmniPro and Compass. The %DE was calculated using a planned dose volume histogram (DVH) created in Monaco's treatment planning system (TPS), which relates radiation dose to tissue and the patient's predicted dose volume histogram in Compass. Statistical correlation between %GP and %DE was verified using Pearson's correlation coefficient. Sensitivity was calculated based on the receiver operating characteristics (ROC) curve. Plans were calculated using Collapsed Cone Convolution and the Monte Carlo algorithm.
Results:
The t-test results of the planned and estimated DVH showed that the mean values were comparable (P > 0.05). For the 3%/3 mm criterion, the average %GP was acceptable for the prostate and endometrial cancer groups, with average rates of 99.68 ± 0.49% and 99.03 ± 0.59% for 2D and 99.86 ± 0.39% and 99.53 ± 0.44% for 3D, respectively. The number of correlations was poor for all analysed data. The mean Pearson's R-values for prostate and endometrial cancer were < 0.45 and < 0.43, respectively. The area under the ROC curve for the prostate and endometrial cancer groups, was lower than 0.667.
Conclusions:
Analysis of the %GP versus %DE values revealed only weak correlations between 2D and 3D verifications. DVH results obtained using the Compass system will be helpful in confirming that the analysed plans respect dosimetric constraints.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
08:34Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
Related Concept Videos
2D NMR: Overview of Homonuclear Correlation Techniques
COSY90 is the standard two-dimensional (2D) COSY experiment that...
2D NMR: Overview of Heteronuclear Correlation Techniques
2D NMR: Homonuclear Correlation Spectroscopy (COSY)
Probability Histograms
Self-Evaluation: Self-Enhancement and Self-Verification
First Pass Effect
Depending on the route of administration, drugs can be metabolized in the liver, intestine, lungs, and vasculature. Orally administered drugs are first absorbed through the...