Performance evaluation of computed tomography systems: Summary of AAPM Task Group 233
Ehsan Samei1, Donovan Bakalyar2, Kirsten L Boedeker3
1Duke University, 2424 Erwin Rd, Durham, NC, 27710, USA.
Background:
The rapid development and complexity of new x-ray computed tomography (CT) technologies and the need for evidence-based optimization of image quality with respect to radiation and contrast media dose call for an updated approach towards CT performance evaluation.
Aims:
This report offers updated testing guidelines for testing CT systems with an enhanced focus on the operational performance including iterative reconstructions and automatic exposure control (AEC) techniques.
Materials And Methods:
The report was developed based on a comprehensive review of best methods and practices in the scientific literature. The detailed methods include the assessment of 1) CT noise (magnitude, texture, nonuniformity, inhomogeneity), 2) resolution (task transfer function under varying conditions and its scalar reflections), 3) task-based performance (detectability, estimability), and 4) AEC performance (spatial, noise, and mA concordance of attenuation and exposure modulation). The methods include varying reconstruction and tube current modulation conditions, standardized testing protocols, and standardized quantities and metrology to facilitate tracking, benchmarking, and quantitative comparisons.
Results:
The methods, implemented in cited publications, are robust to provide a representative reflection of CT system performance as used operationally in a clinical facility. The methods include recommendations for phantoms and phantom image analysis.
Discussion:
In line with the current professional trajectory of the field toward quantitation and operational engagement, the stated methods offer quantitation that is more predictive of clinical performance than specification-based approaches. They can pave the way to approach performance testing of new CT systems not only in terms of acceptance testing (i.e., verifying a device meets predefined specifications), but also system commissioning (i.e., determining how the system can be used most effectively in clinical practice).
Conclusion:
We offer a set of common testing procedures that can be utilized towards the optimal clinical utilization of CT imaging devices, benchmarking across varying systems and times, and a basis to develop future performance-based criteria for CT imaging.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
5-Number Summary
In a box plot, the minimum and maximum data values represent the lower and upper whiskers in the graph, and the median is designated as the center of the box in the chart. The first quartile and third...
Imaging Studies III: Computed Tomography
Discharge Summary Forms
Here's a detailed look at the key components and guidelines for preparing a discharge summary:
Second Order systems II
First Order Systems
When a first-order system is subjected to a unit-step input, its response is characterized by its transfer function. By applying the Laplace transform of the unit-step input to the transfer function, expanding the...