Related Experiment Video
Updated: Jan 20, 2026

Testing Tactile Masking between the Forearms
Published on: February 10, 2016
An Experimental Validation of Masking in IEC 60601-1-8:2006-Compliant Alarm Sounds
Matthew L Bolton1, Xi Zheng1, Meng Li1
1University at Buffalo, The State University of New York, USA.
Objective:
This research investigated whether the psychoacoustics of simultaneous masking, which are integral to a model-checking-based method, previously developed for detecting perceivability problems in alarm configurations, could predict when IEC 60601-1-8-compliant medical alarm sounds are audible.
Background:
The tonal nature of sounds prescribed by IEC 60601-1-8 makes them potentially susceptible to simultaneous masking: where concurrent sounds render one or more inaudible due to human sensory limitations. No work has experimentally assessed whether the psychoacoustics of simultaneous masking accurately predict IEC 60601-1-8 alarm perceivability.
Method:
In two signal detection experiments, 28 nursing students judged whether alarm sounds were present in collections of concurrently sounding standard-compliant tones. The first experiment used alarm sounds with single-frequency (primary harmonic) tones. The second experiment's sounds included the additional, standard-required frequencies (often called subharmonics). T tests compared miss, false alarm, sensitivity, and bias measures between masking and nonmasking conditions and between the two experiments.
Results:
Miss rates were significantly higher and sensitivity was significantly lower for the masking condition than for the nonmasking one. There were no significant differences between the measures of the two experiments.
Conclusion:
These results validate the predictions of the psychoacoustics of simultaneous masking for medical alarms and the masking detection capabilities of our method that relies on them. The results also show that masking of an alarm's primary harmonic is sufficient to make an alarm sound indistinguishable.
Application:
Findings have profound implications for medical alarm design, the international standard, and masking detection methods.
Related Concept Videos
Soundness of Cement
Heart Sounds
Auscultation is the process of listening to these internal body sounds using a stethoscope. The heart produces four types of sounds, but only two—S1 and S2—can usually be heard with a stethoscope.
S1, also known as the "lub" sound, is caused by the closure of atrioventricular (A-V)...
Korotkoff Sounds
During blood pressure assessment, inflating the cuff 30 millimeters of mercury above the patient's systolic blood pressure...
Reliability and Validity
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Masking and Demasking Agents
There are many masking agents, such as cyanide, fluoride, triethanolamine, thiourea, and 2,3-bis(sulfanyl)propan-1-ol (formerly 2,3-dimercapto-1-propanol), with the masking agent chosen based on...