The Supraclavicular Fossa Ultrasound View for Central Venous Catheter Placement and Catheter Change Over Guidewire
Published on: December 23, 2014
Ultrasound to Localize the Peripherally Inserted Central Catheter Tip Position in Newborn Infants
Xiao-Ling Ren1, Hong-Lei Li1, Jing Liu1
1Department of Neonatology and NICU, Beijing Chaoyang District Maternal and Child Healthcare Hospital, Beijing, China.
Objective:
This study aimed to evaluate the application of ultrasound for the localization of the tip position of peripherally inserted central catheters (PICCs) in newborn infants.
Study Design:
This study was a retrospective analysis on ultrasonic localization for PICC placement conducted in our department over the past 2 years. Ultrasonic localization was performed immediately after PICC placement in all neonatal patients. Successful PICC placement was confirmed if the PICC tip position was located at the inferior/superior cavoatrial junction. Chest X-ray localization was performed on 32 infants immediately after ultrasound examination to compare the accuracy of ultrasound localization.
Results:
Of the 186 patients, 174 (93.5%) had successful PICC placement on the first attempt. In 11 (5.9%) patients, the catheter tip was placed beyond the ideal location as follows: too deep (in the right atrium) in 4 patients, too shallow in 4 patients, and malpositioned in 3 patients. Both the sensitivity and the specificity of ultrasound for identifying PICC tip localization were 100%. Complications occurred in 2.7% of this group of patients.
Conclusion:
Ultrasonic localization of the PICC tip position is a timely, accurate, and reliable method and can identify the catheter tip with high accuracy. This method could be widely applied in neonatal wards.
More Related Videos
09:57A Protocol to Set Up Needle-Free Connector with Positive Displacement on Central Venous Catheter in Intensive Care Unit
Published on: July 13, 2019
10:09Drug Treatment by Central Venous Catheter in a Mouse Model of Angiotensin II Induced Abdominal Aortic Aneurysm and Monitoring by 3D Ultrasound
Published on: August 4, 2022
Related Concept Videos
Peripherally and Centrally Acting Muscle Relaxants: A Comparison
Centrally acting muscle relaxants can be further divided into spasmolytic and antispasmodic drugs. Spasmolytic...
The Central Dogma
Measures of Central Tendency
Position-effect Variegation
Serial Position Effect
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):