Unexpectedly large dose rate dependent output from a linear accelerator.

P C Cheng1, H Kubo

  • 1Department of Radiation Oncology, Albany Medical College, New York.

Medical Physics
|September 1, 1988
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Safety and efficacy of guselkumab in Japanese patients with moderate-to-severe plaque psoriasis: a randomized, placebo-controlled, ascending-dose study.

The British journal of dermatology·2017
Same author

SU-E-I-63: Performance Study of An Electron-Tracking Compton Camera for Medical Imaging.

Medical physics·2017
Same author

Note: Optimization of magnifying a polarization angle with Littrow layout blazed gratings.

The Review of scientific instruments·2017
Same author

Design of tangential viewing phase contrast imaging for turbulence measurements in JT-60SA.

The Review of scientific instruments·2016
Same author

HDlive imaging of meconium peritonitis.

Ultrasound in obstetrics & gynecology : the official journal of the International Society of Ultrasound in Obstetrics and Gynecology·2015
Same author

Measurement of the inclusive electron neutrino charged current cross section on carbon with the T2K near detector.

Physical review letters·2014
Same journal

Correction to "On the shape of the radiation survival curve in tumor spheroids: The role of oxygen heterogeneity".

Medical physics·2026
Same journal

Multi-view constrained semi-supervised vertebra detection for 3D ultrasound spine volume.

Medical physics·2026
Same journal

Accuracy of quantitative <sup>177</sup>Lu SPECT/CT imaging: A systematic review.

Medical physics·2026
Same journal

Physics-constrained dual-domain network for CBCT reconstruction from orthogonal X-rays in gynecologic radiotherapy.

Medical physics·2026
Same journal

Decomposition-based harmonization for quantitative PET imaging across scanners and radiotracers.

Medical physics·2026
Same journal

Development and evaluation of an in vivo dose-based monitoring system for electron FLASH radiation therapy.

Medical physics·2026
See all related articles

Routine calibration of a Varian linear accelerator revealed that absorbed dose decreased with increasing dose rate. A faulty integrating board circuit caused up to a 20% difference in absorbed dose for electron beams, impacting radiation therapy accuracy.

Area of Science:

  • Medical Physics
  • Radiation Oncology
  • Electronic Engineering

Background:

  • Accurate calibration of linear accelerators is crucial for effective radiation therapy.
  • Variations in absorbed dose can significantly impact treatment efficacy and patient safety.
  • Routine calibration of Varian Clinac-20 linear accelerators is a standard quality assurance procedure.

Purpose of the Study:

  • To identify the cause of a decreasing absorbed dose with increasing dose rate during linear accelerator calibration.
  • To analyze the electronic failure responsible for dose discrepancies.
  • To propose a method for detecting similar failures to improve quality assurance.

Main Methods:

  • Performed routine calibration of a Varian Clinac-20 linear accelerator.

Related Experiment Videos

  • Investigated dose output variations across different dose rates for a 20-MeV electron beam.
  • Analyzed the electronic circuit of the integrating board suspected of failure.
  • Main Results:

    • A decrease in absorbed dose was observed as the dose rate increased.
    • Up to a 20% difference in absorbed dose was recorded between 100 and 500 monitor units per minute for a 20-MeV electron beam.
    • The root cause was identified as a failure in the electronics circuit of an integrating board.

    Conclusions:

    • Electronic failures in integrating boards can lead to significant absorbed dose inaccuracies in linear accelerators.
    • Early detection and isolation of such failures are essential for maintaining radiation therapy precision.
    • This study provides insights for physicists and engineers to identify and prevent similar calibration issues.