Concurrent validity and intratester reliability of the video-based system for measuring gait poststroke
Nilar Aung1,2, Sunee Bovonsunthonchai1, Vimonwan Hiengkaew1
1Faculty of Physical Therapy, Mahidol University, Nakhon Pathom, Thailand.
Background And Purpose:
Spatio-temporal parameters are commonly used in gait assessment. Advanced tools provide valid and reliable data, considered very effective for physiotherapy intervention. However, these tools may be limited in clinical usage caused by complicated applicability, inaccessibility, and high cost. Therefore, a video-based system is an alternative choice that is easy and affordable for the clinical setting. The purpose of the study was to evaluate the concurrent validity of the video-based system against the validated instrumented gait system (Force Distribution Measurement [FDM]) on the spatio-temporal gait parameters in individuals with stroke. In addition, the intratester reliability of a novice tester was determined.
Methods:
Twenty individuals with stroke participated in the study. Gait was captured by the video-based and FDM systems simultaneously to measure the degree of concurrent validity. Parameters composed of the affected and unaffected step lengths (cm) and step time (s), stride length (cm), gait velocity (m/s), and cadence (steps/min). Pearson correlation coefficient, paired t test, and intraclass correlation coefficient (ICC) were used to determine the concurrent validity, the difference of the data, and intratester reliability.
Results:
All spatio-temporal gait parameters showed excellent degrees of correlation (rp = .94 to.99, p <.001) between the video-based and FDM systems. No significant difference in all parameters was found between the two systems. Excellent intratester reliability (ICC3,1 = 0.91 to 0.99, p < .001) of all gait parameters were found in a novice tester.
Conclusion:
The video-based system was valid and reliable for a novice tester to measure the spatio-temporal gait parameters in individuals with stroke.
Related Concept Videos
Reliability and Validity
Distribution Reliability and Automation
Data Validation
Nursing assessment guides are generally based on holistic models rather than medical...
Data Validation
Key parameters for method validation include:
Formulating and Validating Nursing Diagnosis II
Risk nursing diagnoses represent clinical judgments of an individual, family, or community more vulnerable to developing the health problem than others...
Formulating and Validating Nursing Diagnosis I
There are thirteen domains...