Technical Note: Experimental characterization of the dose deposition in parallel MRI-linacs at various magnetic field
Jarrad Begg1,2,3, Sarah J Alnaghy2,4, Trent Causer2,4
1Department of Medical Physics, Liverpool and Macarthur Cancer Therapy Centre, Liverpool, NSW, 2170, Australia.
Purpose:
Dose deposition measurements for parallel MRI-linacs have previously only shown comparisons between 0 T and a single available magnetic field. The Australian MRI-Linac consists of a magnet coupled with a dual energy linear accelerator and a 120 leaf Multi-Leaf Collimator with the radiation beam parallel to the magnetic field. Two different magnets, with field strengths of 1 and 1.5 T, were used during prototyping. This work aims to characterize the impact of the magnetic field at 1 and 1.5 T on dose deposition, possible by comparing dosimetry measured at both magnetic field strengths to measurements without the magnetic field.
Methods:
Dose deposition measurements focused on a comparison of beam quality (TPR20/10 ), PDD, profiles at various depths, surface doses, and field size output factors. Measurements were acquired at 0, 1, and 1.5 T. Beam quality was measured using an ion chamber in solid water at isocenter with appropriate TPR20/10 buildup. PDDs and profiles were acquired via EBT3 film placed in solid water either parallel or perpendicular to the radiation beam. Films at surface were used to determine surface dose. Output factors were measured in solid water using an ion chamber at isocenter with 10 cm solid water buildup.
Results:
Beam quality was within ±0.5% of the 0 T value for the 1 and 1.5 T magnetic field strengths. PDDs and profiles showed agreement for the three magnetic field strengths at depths beyond 20 mm. Deposited dose increased at shallower depths due to electron focusing. Output factors showed agreement within 1%.
Conclusion:
Dose deposition at depth for a parallel MRI-linac was not significantly impacted by either a 1 or 1.5 T magnetic field. PDDs and profiles at shallow depths and surface dose measurements showed significant differences between 0, 1, and 1.5 T due to electron focusing.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:43A Non-invasive and Technically Non-intensive Method for Induction and Phenotyping of Experimental Bacterial Pneumonia in Mice
Published on: September 28, 2016
Related Concept Videos
Magnetic Fields
A magnetic field is defined by the force that a charged particle experiences...
Magnetic Force Between Two Parallel Currents
The force exerted by the magnetic field due to the first conductor over a finite length of the second conductor is given as the product of the current in the second conductor and the vector product of the length vector along the current element and the field due to the first conductor. According to the...
Electric Field of Parallel Conducting Plates
Consider a cross-section of a thin, infinite conducting plate having a positive charge. For such a large thin plate, as the thickness of the plate tends to zero, the positive charges lie on the plate's two large faces. Without an external electric field, the...
Magnetic Field of a Solenoid
Consider a solenoid with 100 turns wrapped around a cylinder of...
Magnetic Field Lines
Magnetic field lines follow several hard-and-fast rules:
Energy In A Magnetic Field
Take an ideal inductor with zero resistance. Although it's practically impossible, assume that the coil's resistance is so small that it is practically negligible. The loss of the field's energy to dissipate thermal energy (or heat) is thus...