A cost-effective approach to factor assay calibration using a truncated live calibration curve
Susan Guy1, Anne M Sermon-Cadd2, Fiona M Shepherd1
1Sheffield Haemophilia and Thrombosis Centre, Royal Hallamshire Hospital, Sheffield, UK.
Introduction:
The measurements of clotting factor activities are usually performed using a one-stage clotting assay (OSA) or chromogenic substrate assay (CSA). Advances in automated coagulation analysers have led to the utilization of stored calibration curves. There are sometimes substantial intervals between test calibration and analysis of samples. Variability in results can be influenced by calibrant and methodology. Several guidelines recommend calibration and patient samples be performed together in parallel; this incurs costs, but reliance on a stored calibration curve may lead to inaccuracy of results over time.
Methods:
We evaluated inclusion of a live truncated (3 point) calibration curve using calibrator plasma alongside test samples and compared results calculated against the stored calibration curves and live truncated calibration to assess the impact on precision and accuracy. The feasibility of this was tested on two hospital sites in the UK; OSA and CSA were performed on Sysmex CS5100 using Actin FS, SynthASil, Innovin, Biophen chromogenic VIII and Rossix chromogenic IX.
Results:
Results of two batches of IQC were compared for FII, FV, FVII, FX, FIX:C, FXI, FXII and OSA FVIII (FVIII:C1) and CSA FVIII:C (FVIII:CR) and FXI:C. By utilizing a live truncated calibration, precision improved with the most striking examples: FXII %CV 23.1% to 3.1% (site A) FXI 7.3% to 2.4% (site B). The improvement in other clotting factors was more modest.
Conclusion:
To the best of our knowledge, this is the first study that demonstrates that the use of a live truncated calibration curve will improve precision and accuracy.
Related Concept Videos
Calibration Curves: Correlation Coefficient
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Plotting and Calibrating the Root Locus
The maximum gain occurs at the breakaway points between open-loop poles on the real axis, while the minimum gain is...
Truncation in Survival Analysis
Left truncation occurs when individuals who experienced the event of interest before a certain time are not included in the study. This is often due to a "delayed entry" into the study where only those who survive until a certain entry point are...