Evaluation of right ventricular volume and systolic function in normal fetuses using intelligent spatiotemporal image
Jia-Xing Sun1, Ai-Lu Cai1, Li-Mei Xie1
1Ultrasound Department, Shengjing Hospital of China Medical University, Shenyang 110004, Liaoning Province, China.
Background:
Heart defects are the most common congenital malformations in fetuses. Fetal cardiac structure and function abnormalities lead to changes in ventricular volume. As ventricular volume is an important index for evaluating fetal cardiovascular development, an effective and reliable method for measuring fetal ventricular volume and cardiac function is necessary for accurate ultrasonic diagnosis and effective clinical treatment. The new intelligent spatiotemporal image correlation (iSTIC) technology acquires high-resolution volumetric images. In this study, the iSTIC technique was used to measure right ventricular volume and to evaluate right ventricular systolic function to provide a more accurate and convenient evaluation of fetal heart function.
Aim:
To investigate the value of iSTIC in evaluating right ventricular volume and systolic function in normal fetuses.
Methods:
Between October 2014 and September 2015, a total of 123 pregnant women received prenatal ultrasound examinations in our hospital. iSTIC technology was used to acquire the entire fetal cardiac volume with off-line analysis using QLAB software. Cardiac systolic and diastolic phases were defined by opening of the atrioventricular valve and the subsequent closure of the atrioventricular valve. The volumetric data of the two phases were measured by manual tracking and summation of multiple slices and recording of the right ventricular end-systolic volume and the right ventricular end-diastolic volume. The data were used to calculate the right stroke volume, the right cardiac output, and the right ejection fraction. The correlations of changes between the above-mentioned indices and gestational age were analyzed. The right ventricular volumes of 30 randomly selected cases were measured twice by the same sonographer, and the intra-observer agreement measurements were calculated.
Results:
Among the 123 normal fetuses, the mean right ventricular end-diastolic volume increased from 0.99 ± 0.34 mL at 22 wk gestation to 3.69 ± 0.36 mL at 35+6 wk gestation. The mean right ventricular end-systolic volume increased from 0.43 ± 0.18 mL at 22 wk gestation to 1.36 ± 0.22 mL at 35+6 wk gestation. The mean right stroke volume increased from 0.62 ± 0.29 mL at 22 wk gestation to 2.33 ± 0.18 mL at 35+6 wk gestation. The mean right cardiac output increased from 92.23 ± 40.67 mL/min at 22 wk gestation to 335.83 ± 32.75 mL/min at 35+6 wk gestation. Right ventricular end-diastolic volume, right ventricular end-systolic volume, right stroke volume, and right cardiac output all increased with gestational age and the correlations were linear (P < 0.01). Right ejection fraction had no apparent correlation with gestational age (P > 0.05).
Conclusion:
Fetal right ventricular volume can be quantitatively measured using iSTIC technology with relative ease and high repeatability. iSTIC technology is expected to provide a new method for clinical evaluation of fetal cardiac function.
More Related Videos
Related Concept Videos
Correlations
Correlation and Causation
Correlation versus Causation
If the dependent variable increases or decreases when the independent variable increases, there is a positive or negative...
Intelligence
Development of the Sexual Organs in the Embryo and Fetus
Near the gonadal ridges, two duct systems are present: the mesonephric ducts (Wolffian ducts) and paramesonephric ducts (Müllerian ducts). These ducts form the basis for the...
Measures of Intelligence
Validity refers to how well a test measures what it claims to measure. An intelligence test should accurately assess intelligence rather than another characteristic, like anxiety. Criterion validity is one way to evaluate this;...
Correlation
Two variables, for example, a and b, are said to be positively correlated if both variables move in the same direction. In other words, a positive correlation exists between two variables, a and b, if: