Pulmonary vein isolation using a higher power shorter duration CLOSE protocol with a surround flow ablation catheter
Benjamin Berte1, Gabi Hilfiker1, Ian Russi1
1Heart Center, Luzerner Kantonsspital, Lucerne, Switzerland.
Introduction:
The CLOSE protocol combines ablation index (AI) and ≤6 mm interlesion distance using standard power settings for the treatment of atrial fibrillation (AF). The purpose was to compare the safety and efficacy of a conventional CLOSE and a higher power shorter duration (HPSD)-CLOSE pulmonary vein isolation (PVI) strategy.
Methods And Results:
All consecutive patients referred for PVI were included after informed consent was obtained from them. Group 1 was treated with a standard CLOSE protocol and group 2 with a HPSD-CLOSE protocol (45 W anterior and 35 W posterior). Procedural parameters and 6-month follow-up were analyzed. In total, 174 patients (group 1: n = 94 [paroxysmal: n = 74]; group 2: n = 80 [paroxysmal: n = 65], similar baseline characteristics) were included. PVI was reached in all, but procedure duration (82 ± 18 minutes vs 100 ± 22 minutes; P < .0001) and radiofrequency (RF) time (23 ± 5 minutes vs 36 ± 11 minutes; P < .0001) was shorter in group 2. First pass isolation was similar in groups 2 and 1 (left veins: 94% vs 90%; P = .42 and right veins: 83% vs 84%; P = .79, respectively). Six-month off- antiarrhythmic drugs freedom of AF/AT was similar in groups 2 and 1 (82% [paroxysmal: 86%] vs 83% [paroxysmal: 88%]; P = .93, respectively). Major complications were similar (group 2: 1% vs group 1: 3%; P = .39).
Conclusion:
A higher-power short duration approach can shorten a CLOSE procedure and reduce ablation time without having a negative impact on safety or efficiency.
More Related Videos
11:03Direct Pressure Monitoring Accurately Predicts Pulmonary Vein Occlusion During Cryoballoon Ablation
Published on: February 26, 2013
08:00Reduction of Iatrogenic Atrial Septal Defects with an Anterior and Inferior Transseptal Puncture Site when Operating the Cryoballoon Ablation Catheter
Published on: June 15, 2015
Related Concept Videos
The Power Flow Problem and Solution
Control of Power Flow
Maximum Power Flow and Line Loadability
Power
Veins
Structure of Veins:
The structure of veins is specifically designed to assist in the low-pressure transportation of...
Close Relationships and Culture