Ultrasound-based repositioning and real-time monitoring for abdominal SBRT in DIBH
Judit Boda-Heggemann1, Dwi Seno Kuncoro Sihono2, Lara Streb1
1University Medical Center Mannheim, Department of Radiation Oncology, University of Heidelberg, Germany.
Aim:
Ultrasound-based repositioning and real-time-monitoring aim at the improvement of the precision of SBRT in deep inspiration breath-hold (DIBH). Accuracy of ultrasound-based daily repositioning was estimated by comparison with DIBH-cone-beam-CT. Intrafraction motion during beam-delivery was assessed by ultrasound-real-time-monitoring.
Patients/Methods:
Residual error after ultrasound-based interfractional repositioning (85 fractions, 16 SBRT-series; 14 patients) was assessed by marker-based (7 series) or liver-contour-based (9 series) matching in DIBH-CBCT. During beam-delivery, the percentage of 3D misalignment vector below 2 mm, between 2 and 5 mm, 5-7 mm and over 7 mm was estimated. Percentage of relevant target-displacements was analyzed as a function of DIBH-duration.
Results:
Residual error after ultrasound-based positioning was 0.4 ± 3.3 mm in LR (left-right), 0.2 ± 4.3 mm in CC (cranio-caudal) and 1.0 ± 3.0 mm in AP (anterior-posterior) directions (vector magnitude 5.4 ± 3.3 mm, MV ± SD). Over 544 DIBHs, target displacement was 1.3 ± 0.5 mm, 0.7 ± 0.3 mm, 1.6 ± 0.6 mm for CC, LR and AP directions, respectively (3D-vector 2.5 ± 0.7 mm). 3D misalignment vector length was below 2 mm in 49.8%, between 2 and 7 mm in 46.3%, and over 7 mm in 3.9% of the beam-delivery-time. During the first 5 s of the DIBH, 3D-misalignment vector length was always below 10 mm. Percentage of target displacements over 10 mm was 0.2%, 0.5% and 0.8% for 10 s, 15 s and 20 s DIBH-duration.
Conclusions:
Ultrasound-based interfractional repositioning is an accurate method for daily localization of abdominal DIBH-SBRT targets. Residual motion is <7 mm in 96% of the beam-delivery-time. Deviations >10 mm occur rarely and can be avoided by gating the beam at a predefined threshold. Ideal DIBH-duration should not exceed 15 s.
Related Concept Videos
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...
Real Time RT-PCR
The real-time quantification of the number of amplified products is...
Abdominal Aorta
The celiac trunk, a singular artery, divides into the left gastric artery, which...
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Abdominal Regions and Quadrants
The simpler quadrants approach, which is more commonly used in medicine, subdivides the cavity with one horizontal and one vertical line that intersects at the patient's umbilicus (navel). The four...
Self-Presentation: Self-Monitoring and Self-Handicapping