Validation of posterior corneal curvature measurements with color light-emitting diode topography
David P Piñero1,2, Vicent J Camps1, Dolores de Fez1
1Department of Optics, Pharmacology and Anatomy, University of Alicante, Alicante, Spain.
Purpose:
To evaluate the intrasession repeatability and validity of posterior corneal curvature and astigmatism measurements provided by a color light-emitting diode reflection topography system in healthy eyes.
Methods:
A total of 40 healthy eyes of 40 patients (age, 16-66 years) were enrolled. A complete eye examination was performed in all cases including posterior topographic analysis with two systems: the Scheimpflug-based system (Pentacam; Oculus Optikgeräte GmbH, Wetzlar, Germany) and the Cassini system (i-Optics; Ophthec, The Hague, The Netherlands). With this last system, three consecutive measurements were taken to assess the level of intrasession repeatability (within-subject standard deviation, Sw; intraclass correlation coefficient). The Bland & Altman analysis was used to evaluate the interchangeability of both devices.
Results:
The Sw was ⩽0.06 mm for all posterior corneal radius measurements, with intraclass correlation coefficient of ⩾0.960. The Sw for the magnitude of astigmatism, J0, and J45 were 0.15, 0.04, and 0.04 D, respectively, with intraclass correlation coefficient values of 0.876, 0.897, and 0.840, respectively. Statistically significant differences between devices were found in all parameters evaluated (p ⩽ 0.025). The interchangeability analysis revealed the presence of clinically relevant limits of agreement for the flattest (0.03 to 0.50 mm) and steepest posterior corneal radii (-0.01 to 0.39 mm). In contrast, limits of agreements were not clinically relevant for the magnitude of posterior astigmatism (-0.17 to 0.27 D) and their power vector components (-0.11 to 0.15 D).
Conclusion:
The Cassini system provides consistent measures of posterior corneal curvature and astigmatism in healthy eyes, but only measures of posterior astigmatism can be considered as interchangeable with those provided by the Pentacam.
More Related Videos
09:03Optimizing Photoneuromodulation Techniques to Evaluate the Role of Green Light-Emitting Diodes in Pain Management
Published on: March 28, 2025
10:41Enhanced Electron Injection and Exciton Confinement for Pure Blue Quantum-Dot Light-Emitting Diodes by Introducing Partially Oxidized Aluminum Cathode
Published on: May 31, 2018
Related Concept Videos
Degree of Curvature and Radius of Curvature
Methods of Obtaining Topography
Colors and Magnetism
When atoms or molecules absorb light at the proper frequency, their electrons are excited to higher-energy orbitals. For many main group atoms and molecules, the absorbed photons are in the ultraviolet range of the electromagnetic spectrum, which cannot be detected by the human eye. For coordination compounds, the energy difference between the d orbitals often allows photons in the visible range to be absorbed and emitted, which is seen as colors by the human...
Reliability and Validity
Vertebral Column: Regions and Curvature
Regions of the Vertebral Column
In an adult, the spine is subdivided into five regions: the cervical, the thoracic, the lumbar, the sacral, and the coccygeal region. The spine initially develops as a series of 33 vertebrae; after 20 years of age, the nine bones in the sacral region, five sacral, and four coccygeal bones fuse to form...
Zener Diodes