Accuracy of Using a Point-of-Care Glucometer for Cerebrospinal Fluid Glucose Screening in Resource-Limited Countries
Ratchada Kitsommart1, Thananjit Wongsinin2,3, Uraporn Swasee2
1Division of Neonatology, Department of Pediatrics, Faculty of Medicine Siriraj Hospital, Mahidol University, Bangkok, Thailand; rkitsommart@hotmail.com.
Objectives:
To explore the accuracy of using a point-of-care (POC) glucometer for cerebrospinal fluid (CSF) glucose screening.
Methods:
A cross-sectional study was conducted. A glucose analysis of CSF samples collected from infants <90 days with suspected meningitis was paired between tests by using a POC glucometer (POC-CSF glucose) and a laboratory glucose analysis (laboratory-CSF glucose). Accuracy and limits of agreement were compared, as well as the glucometer performance to detect a laboratory-CSF glucose level <45 and 60 mg/dL.
Results:
Seventy-three CSF samples were analyzed. Subjects' mean gestational age was 32.2 (SD 4.0) weeks, the mean weight was 1947.7 (SD 814.5) g, and the median age was 8 (interquartile range: 2 to 19.5) days. POC-CSF glucose levels ranged from 26 to 126 mg/dL. The mean (±1.96 SD) difference between POC-CSF and laboratory-CSF glucose levels was -1.6 (interquartile range: -12.6 to 9.4) mg/dL. A POC-CSF glucose level <45 mg/dL has a sensitivity and negative predictive value (NPV) to detect a laboratory-CSF glucose level <45 mg/dL of 82% and 94%, respectively. For a laboratory-CSF glucose level <60 mg/dL, a POC glucose level <60 mg/dL provides a sensitivity and NPV of 96% and 90%, respectively, whereas sensitivity and NPV reach 100% at a POC glucose level <70 mg/dL.
Conclusions:
A POC glucometer for CSF glucose can detect a potential abnormal glucose level with an appropriate cutoff level. This may facilitate rapid decisions for empirical antibiotics in suspected meningitis, pending laboratory results in limited-resource settings, but requires robust validation in future studies before implementation.
Related Concept Videos
Cerebrospinal Fluid
CSF Production
CSF is produced mainly in the choroid plexus, a network of capillaries and ependymal cells located within the ventricular system of the brain....
Accuracy, limits, and approximation
Accuracy is defined as the closeness of the measured value to the true or actual value. In engineering mechanics, repeated measurements are taken during theoretical or experimental analyses to ensure that the result is precise and accurate.
The accuracy of any solution is based on the...
Improving Translational Accuracy
Short-distance Transport of Resources
Uncertainty in Measurement: Accuracy and Precision
Limiting Reactant