Precontoured Quadrilateral Surface Acetabular Plate Fixation Demonstrates Increased Stability When Compared With
Weston Ryan1, Nicholas A Alfonso2, Todd Baldini2
1University of Colorado School of Medicine, Aurora, CO.
Objective:
To compare the stability of 3 fixation strategies for a transverse acetabular fracture: a reconstruction plate with anterior and posterior column screws (group A); an infrapectineal precontoured quadrilateral surface buttress (iPQSB) plate alone (group B); and an anterior column lag-screw and iPQSB plate (group C).
Methods:
A transverse acetabular fracture was created in 18 synthetic hemipelvises. Six were fixed by each of the 3 methods described. Specimens underwent cyclic axial compressive loading to 1700N for 42,000 cycles while anterior and posterior column displacements were measured, followed 4800N for 50 cycles. Displacement and stiffness data were analyzed with analysis of variance and Tukey HSD. A Cox proportional hazards regression model was used to determine survival rate. P values < 0.05 were considered significant.
Results:
Group C had significantly less posterior column displacement (0.16 ± 0.06 mm) compared with group B (0.38 ± 0.37 mm, P < 0.0001) and group A (0.38 ± 0.37 mm, P < 0.0001). In addition, group A had significantly more anterior column displacement (0.28 ± 0.11 mm) than group B (0.22 ± 0.14 mm, P = 0.0310) and group C (0.18 ± 0.09 mm, P = 0.0001). Group C was 10.5% stiffer than group A (P = 0.0037). Group B had a 7.27x greater rate of failure than group C (95% confidence interval, 1.6-33.2).
Discussion And Conclusion:
Under anatomical loading, iPQSB plates with anterior column lag-screw fixation demonstrate increased stability in a synthetic bone transverse acetabular fracture model. Based on our data, we support additional evaluation of early weight-bearing after transverse acetabular fracture fixation in patients with healthy bone when an anterior column screw-iPQSB plate construct is used.
More Related Videos
Related Concept Videos
Column Efficiency: Plate Theory
A higher number of theoretical plates signifies better column efficiency and improved separation capabilities. Plate height affects bandwidth and separation quality; it is inversely...
Electric Field of Parallel Conducting Plates
Consider a cross-section of a thin, infinite conducting plate having a positive charge. For such a large thin plate, as the thickness of the plate tends to zero, the positive charges lie on the plate's two large faces. Without an external electric field, the...
Fluid Pressure over Curved Plate of Constant Width
Steady, Laminar Flow Between Parallel Plates
Fluid Pressure over Flat Plate of Constant Width
The resultant force...
Fluid Pressure over Flat Plate of Variable Width
The pressure distribution on the plate can be calculated by determining the force that acts on a differential area strip of the plate. Thus, the magnitude of the force is equal to the...